AR FI7000 Скачать руководство пользователя страница 43

 

PL7004 Kit 

Rev F 

27 

Code 

Command 

Detail 

command  strings  are  fully  sent  with  less  than  5  seconds  between 
each character.] 

Ed<termination>  Buffer full error 

Too  many  characters  contained  between  the  Start  Character  and 
the  Carriage  Return  of  a  command.  Message  is  not  processed  by 
the probe kit. [Host should send correct string.] 

Ee<termination>  Invalid command 

first character 

The received command is not a valid command per the command 
list  for  the  probe  kit.  Specifically,  the  first  character  of  the 
command is not one of the valid first characters of any acceptable 
command. [Host to check command] 

Ef<termination>  Invalid command 

additional 
character(s) 

The  additional  parameters  contained  in  a  received  command  are 
not  valid.  Specifically  the  additional  characters,  after  the  first 
character in a command, are not valid additional characters for this 
type of command for the probe kit. [Host to check command.] 

Eg<termination>  Probe hardware 

error 

An error has been detected by limited error detection in the probe 
(not  the  interface)  and  it  has  been  determined  that  this  error  will 
cause  the  requested  response  to  be  unreliable,  so  the  requested 
response is not sent. [Host should try the command again to see if 
this  probe  error  clears.  If  error  does  not  clear  probe  needs  to  be 
repaired.  Note that since  the  probe is  not  user  repairable, there  is 
no reason for further fault detection in the probe.]  

Eh<termination>  Laser not on error 

Commands  cannot  be  sent  to  or  received  from  the  probe  without 
the laser power on.  [Turn laser on from front panel and check for 
fault status.] 

 

3.7.2 

GPIB (IEEE-488) Operating Protocols 

3.7.2.1 

IEEE-488 (GPIB) Communications 

For  IEEE-488  operation,  the  device  address  is  set  using  the  dip  switches  on  the  rear  panel  of  the  product.  
Ensure  that  each  device  connected  to  the  IEEE-488  buss  is  set  to  a  unique  address.  (Resetting  the  address 
requires re-booting the unit). 

Specific IEEE-488 bus commands depend on which software package you are using.  To send commands be 
sure that the Field Monitors address is set properly and that the controller has correctly identified the unit as a 
“listening” device.  

When sending commands to the unit via the GPIB interface, the command must be terminated with an 

EOI

3.7.2.2 

Setting the IEEE-488 (GPIB) Address 

The IEEE-488 device address can be set to any number between 1 and 30.  This selection is made by setting 
Switches 1 to 5 of the rear panel DIP switch to the binary equivalent of the number.  Table 3-4 illustrates this 
switch selection.  Please note that although addresses 0 and 31 can be entered, neither address is valid for this 
device, and therefore should be avoided. 

Содержание FI7000

Страница 1: ...Operating and Service Manual PL7004 Model 10024183 Part Number Serial Number 160 School House Road Souderton PA 18964 215 723 8181 Fax 866 859 0582 www arworld us...

Страница 2: ......

Страница 3: ...NELEC EN 61010 1 Issued 2010 10 01 Ed 3 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Part 1 General Requirements UL 61010 1 Issued 2012 05 11 Ed 3 Safety Req...

Страница 4: ......

Страница 5: ......

Страница 6: ......

Страница 7: ...clearance on all vented sides of the equipment to permit the airflow required for proper ventilation Do not restrict airflow into the equipment by blocking any vents or air intakes Restricting airflow...

Страница 8: ...a assembly may be heavy requiring two persons to lift Use caution when installing or removing unit Follow all equipment setup and lifting instructions specified in this document Ensure connectors are...

Страница 9: ...von 10 2 cm 4 in auf allen bel fteten Seiten des Ger ts ein um eine ausreichende Luftzirkulation zu gew hrleisten Beintr chtigen Sie den Luftstrom des Ger ts nicht indem Sie L ftungs ffnungen oder de...

Страница 10: ...allen in diesem Dokument angegebenen Anweisungen zur Anbringung Stellen Sie sicher dass alle Anschl sse f r den beabsichtigten Betrieb geeignet sind Informationen zu den Anschl ssen erhalten Sie im Be...

Страница 11: ...tilation Ne limitez pas le flux d air allant dans l quipement en bloquant tout vent ou entr e d air La restriction du flux d air peut endommager l quipement causer des coupures intermittentes ou des d...

Страница 12: ...dent lorsque vous installez ou retirez l unit Respectez toutes les instructions concernant l installation et le levage de l quipement indiqu es dans ce document Assurez vous que les connecteurs sont a...

Страница 13: ...is voor goede ventilatie mogelijk te maken Belemmer de luchtstroom in het apparaat niet door ventilatieopeningen of luchtinlaten te blokkeren Het belemmeren van de luchtstroom kan leiden tot schade a...

Страница 14: ...nstellen en optillen van de apparatuur die in dit document worden vermeld Zorg ervoor dat de connectoren geschikt zijn voor de beoogde werking De connectoren worden gespecificeerd in de gebruikershand...

Страница 15: ...11 2 3 5 Control Logic 11 3 OPERATION 13 3 1 Warnings and Cautions 13 3 2 Laser Safety 13 3 2 1 Safe Operating Procedures and Precautions 13 3 3 General Guidelines For Use of Field Probes 14 3 3 1 De...

Страница 16: ...Upgrade Policies 33 Appendix A Dimensional Drawings 35 FIGURES 1 1 PL7004 1 1 2 FI7000 Interface 1 3 1 PL7004 Probe Features 17 3 2 Interface Assembly Front Panel Features 18 3 3 Probe Interface Asse...

Страница 17: ...ed interface via supplied fiber optic cables and bulkhead feed through adapters The interface provides standard communication protocols RS 232 GPIB USB 2 0 to the user s host computer to relay the mea...

Страница 18: ...accessories for use with this probe kit PS1000 probe stand non conductive FM7004 Field Monitor Long fiber optic cables for single length up to 100 m Bulkhead fiber optic cable feed through adapters s...

Страница 19: ...d for local monitoring and control Features Specifications PL7004 Kit Peak Measuring Electric Field Probe Kit for Pulsed Fields 800MHz 3 6GHz 80 800 V m User selectable X Y Z Axes AR RF Microwave Inst...

Страница 20: ......

Страница 21: ...vide consistent readings for pulse rates as low as 200Hz approximately 1 second after an input change This probe requires an FI7000 for power and communication FM7004A is recommended for local monitor...

Страница 22: ...ied 1 5 dB 1 GHz With correction factors applied Typical expand ed measurement uncertainty 95 confidence interval 0 8 dB 800 MHz 1 GHz 1 4 dB 1GHz 3 6 GHz Response Time Sampling Rate through FI7000 20...

Страница 23: ...ld us Approved for public release by AR RF Microwave Instrumentation 020816 Interfaces IEEE 488 GPIB USB 2 0 Test and Measurement class RS 232 19200 Baud Fiber Optic Serial 19200 Baud reserved for con...

Страница 24: ......

Страница 25: ...ntains the serial number of the probe the revision of the probe s firmware and the probe s most recent calibration date The probe interface sits between the probe and the host and provides all of the...

Страница 26: ...trols the collection of 12 bit A D samples from each input in a rapid sequence repeated for 16 samples of each input and completed in 1200 microseconds This is sufficiently short in time that the read...

Страница 27: ...l Logic The control logic provides operational control and routing of signals It includes a microcontroller It functions to route the host commands received from one specific controller host interface...

Страница 28: ...PL7004 Kit 12 Rev F...

Страница 29: ...WARNING This Class 1 laser product contains an embedded Class 4 laser light source which may be hazardous Only trained qualified personnel should operate maintain and service it WARNING Explosive Atmo...

Страница 30: ...time at which a measurement of that field would be available to be read at a value within about 2 of final value This would be about four time constants of the detector amplifier response plus the tot...

Страница 31: ...ientation to a constant field 3 3 2 Guidelines for use of Field Probes 3 3 2 1 Probe Support Structures It is very important to keep conductive objects away from a field probe Any such objects in the...

Страница 32: ...power in the fundamental to produce a reading Reading square root of sum of squares of the two field levels At higher field levels the detectors may not be in their square law range and tend to respon...

Страница 33: ...4 5 6 7 Figure 3 1 PL7004 Probe Features Table 3 1 PL7004 Probe Features Label Title Function 1 Sensor elements Three antennas sensor elements to sense electric field Each includes one diode detector...

Страница 34: ...e kit and illuminates to indicate that prime power is applied 2 Key Switch Enables the laser power supply 3 Laser On Switch Illuminated momentary switch to start laser on sequence Switch is illuminate...

Страница 35: ...s for future use 2 Open the carrying case which holds all smaller items 3 Check all materials against the packing list to verify that the equipment received matches the order Note any discrepancies If...

Страница 36: ...he fiber cable is rugged but sharp bends less than 1 inch radius must be avoided Position the fiber so that it is not subject to physical abuse and will not be walked upon Do not put excessive stress...

Страница 37: ...F O Status LED Enable the laser power supply with the key switch Press the laser on button once Wait three seconds Observe that the green Laser on indicator stays lit This indicates that the probe is...

Страница 38: ...fer Protocol The probe kit only responds to commands from another device It transmits no data without first receiving instructions to do so All characters are defined by and transferred as common 7 bi...

Страница 39: ...lus one floating decimal point after the 2nd or 3rd digit in implied units of Volts meter and s is a status flag S or X where S indicates status is OK X indicates status is not OK See NOTE 1 Described...

Страница 40: ...CII decimal point 5 characters s Status flag S or X 1 character termination Depends on the selected response termination characters see TERM command for details 1 or 2 characters Character field separ...

Страница 41: ...tion 1 6 characters r Firmware revision level 0 10 characters termination Depends on the selected response termination characters see TERM command for details 1 or 2 characters Total number of charact...

Страница 42: ...e probe kit will generally check and then clear its input buffer until a new start character is received Specific responses are as listed below Recommended actions are shown in brackets termination de...

Страница 43: ...error clears If error does not clear probe needs to be repaired Note that since the probe is not user repairable there is no reason for further fault detection in the probe Eh termination Laser not o...

Страница 44: ...ns with this port requires the host computer to have a USBTMC driver available All commands and queries through this port must be terminated with a CR The cable required to make this connection is a U...

Страница 45: ...peration Command Response Uses the Normal Operation Command Response as defined for RS 232 3 7 3 2 Error Code Responses Uses the Error Code Responses as defined for RS 232 3 7 4 F O Serial Dedicated p...

Страница 46: ...PL7004 Kit 30 Rev F...

Страница 47: ...n the event that the kit malfunctions while it is still under warranty always contact an authorized service representative Contact AR Customer Service for return authorization before returning any par...

Страница 48: ...problem provide the date s identify the contact and the nature of the contact Include the serial number of the item being returned along with the Return Authorization RMA number obtained from the AR...

Страница 49: ...ing within specifications by performing a subset of testing equivalent to the test performed on this unit when shipped This test subset may be changed by AR depending upon the calibration performance...

Страница 50: ...PL7004 Kit 34 Rev F...

Страница 51: ...Rev F 35 Appendix A Dimensional Drawings Figure A 1 Model PL7004...

Страница 52: ...PL7004 Kit 36 Rev F...

Страница 53: ...cover If the security label is removed and or cut the warranty of the module will be voided Vacuum tubes in the L series amplifiers traveling wave tubes in TWT amplifiers and power heads will be free...

Страница 54: ......

Отзывы: