![AR 100A400AM20 Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/ar/100a400am20/100a400am20_operating-and-service-manual_2962301044.webp)
Model 100A400AM20
28
Rev A
4.3.8
Low or No Power Output (RF Test) (Schematic 10047796)
NOTE: The DC Tests specified in Section 4.3.7 should be completed before conducting the RF tests specified
in the following sections.
1.
The Model 100A400AM20’s saturated power and typical gain response at -20 dBm input is shown in
Figure 4-1. The actual gain may vary from that shown in Figure 4-
1 but should be ≥
50 dBm.
Figure 4-1. Saturated Power and Typical Response at -20 dBm Input