Electromagnetic Compatibility:
Electromagnetic Compatibility Information According to IEC/EN 60601-1-2
Table 1 for Emissions
Guidance and Manufacturer’s Declaration –
Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions CISPR 11
Group 1
The CCT1500 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The CCT1500 is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
EN 61000-3-2
Class A
Voltage
Fluctuations/Flicker
Emissions
EN 61000-3-3
Compliant
Table 2 for Electromagnetic Immunity
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 2, 4, 8, 15 kV air
± 6 kV contact
± 2, 4, 8, 15 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/ output
lines
± 2 kV for power
supply lines
± 1 kV for input/ output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
) for
0.5 cycle
40%
U
T
(60% dip in
U
T
) for 5
cycles
70%
U
T
(30% dip in
U
T
) for
25 cycles
<5%
U
T
(>95% dip in
U
T
) for 5
s
<5%
U
T
(>95% dip in
U
T
) for 0.5
cycle
40%
U
T
(60% dip in
U
T
) for 5
cycles
70%
U
T
(30% dip in
U
T
) for
25 cycles
<5%
U
T
(>95% dip in
U
T
) for 5 s
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the CCT1500 requires continued operation during
power mains interruptions, it is recommended that
the CCT1500 be
powered from an uninterruptible
power supply or a battery.
Power frequency (50/60
Hz)
magnetic field
EN 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Conducted RF
EN 61000-4-6
Radiated RF
EN 6100-4-3
3Vrms
0.15MHz to 80 MHz
3V/m
80 MHz
–
2.7 GHz
3Vrms
0.15MHz to 80 MHz
3V/m
80 MHz
–
2.7 GHz
Portable and mobile RF
communications equipment should
be used no closer to any part of the
CCT1500 system, including
cables, than the recommended
separation distance calculated from
the equation applicable to the
frequency of the transmitter.