background image

Electromagnetic Compatibility:

 

Electromagnetic Compatibility Information According to IEC/EN 60601-1-2 

 

Table 1 for Emissions 

Guidance and Manufacturer’s Declaration –

 Electromagnetic Emissions

 

Emissions Test 

Compliance 

Electromagnetic Environment - Guidance 

RF emissions CISPR 11

 

Group 1

 

The CCT1500 uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not likely to 

cause any interference in nearby electronic equipment.

 

RF emissions CISPR 11

 

Class B

 

The CCT1500 is suitable for use in all establishments, including 

domestic establishments and those directly connected to the 

public low-voltage power supply network that supplies 

buildings used for domestic purposes.

 

Harmonic Emissions  
EN 61000-3-2

 

Class A

 

Voltage 

Fluctuations/Flicker 

Emissions 
EN 61000-3-3

 

Compliant

 

 

Table 2 for Electromagnetic Immunity 

Guidance and Manufacturer’s Declaration –

 Electromagnetic Immunity 

Immunity Test 

IEC 60601 Test Level 

Compliance Level 

Electromagnetic Environment - Guidance 

Electrostatic discharge 

(ESD)  
IEC 61000-4-2

 

± 6 kV contact

 

± 2, 4, 8, 15 kV air

 

± 6 kV contact

 

± 2, 4, 8, 15 kV air

 

Floors should be wood, concrete or ceramic tile. If 

floors are covered with synthetic material, the relative 

humidity should be at least 30%.

 

Electrical fast 

transient/burst  
IEC 61000-4-4

 

± 2 kV for power 

supply lines

 

± 1 kV for input/ output 

lines

 

± 2 kV for power 

supply lines

 

± 1 kV for input/ output 

lines

 

Mains power quality should be that of a typical 

commercial or hospital environment.

 

Surge

 

IEC 61000-4-5

 

± 1 kV line(s) to line(s)

 

± 2 kV line(s) to earth

 

± 1 kV line(s) to line(s)

 

± 2 kV line(s) to earth

 

Mains power quality should be that of a typical 

commercial or hospital environment.

 

Voltage dips, short 

interruptions and voltage 

variations on power supply 

input lines

 

IEC 61000-4-11

 

<5% 

U

T

 

(>95% dip in 

U

T

) for 

0.5 cycle

 

40% 

U

T

 

(60% dip in 

U

T

) for 5 

cycles

 

70% 

U

T

 

(30% dip in 

U

T

) for 

25 cycles

 

<5% 

U

T

 

(>95% dip in 

U

T

) for 5 

s

 

<5% 

U

T

 

(>95% dip in 

U

T

) for 0.5 

cycle

 

40% 

U

T

 

(60% dip in 

U

T

) for 5 

cycles

 

70% 

U

T

 

(30% dip in 

U

T

) for 

25 cycles

 

<5% 

U

T

 

(>95% dip in 

U

T

) for 5 s

 

Mains power quality should be that of a typical 

commercial or hospital environment. If the user of 

the CCT1500 requires continued operation during 

power mains interruptions, it is recommended that 

the CCT1500 be

 

powered from an uninterruptible 

power supply or a battery.

 

Power frequency (50/60 

Hz)

 

magnetic field

 

EN 61000-4-8

 

3 A/m

 

3 A/m

 

Power frequency magnetic fields should be at levels 

characteristic of a typical location in a typical 

commercial or hospital environment.

 

Conducted RF 
EN 61000-4-6 
 
Radiated RF 
EN 6100-4-3 

3Vrms 
0.15MHz to 80 MHz 
 
3V/m 
80 MHz 

 2.7 GHz 

3Vrms 
0.15MHz to 80 MHz 
 
3V/m 
80 MHz 

 2.7 GHz 

Portable and mobile RF 
communications equipment should 
be used no closer to any part of the 
CCT1500 system, including 
cables, than the recommended 
separation distance calculated from 
the equation applicable to the 
frequency of the transmitter. 

Содержание CCT1500

Страница 1: ...CCT1500 Cryo Compression Therapy Set Operator s Manual ...

Страница 2: ...ystem 15 Maintaining the CCT 1500 System 16 Troubleshooting 16 General Cautions Warnings 17 Indications for Use and Contraindications 18 Warranty Information 19 Icon Meaning Read instructions before use Type BF Components Double insulated product Direct current power Name and location of product manufacturer Dispose of electrical and electronic components appropriately Warning symbol to indicate p...

Страница 3: ...herapy the CCT1500 Control Unit and Compression Boots provide athletes with the flexibility to use either the cryo therapy or compression therapy functionalities independently This flexibility combined with program customization options delivers the ultimate personalized recovery solution for athletes With the CCT1500 athletes can choose how they recover as well as where they recover The Control U...

Страница 4: ...PA2 AC DC Plug AQU624 AC2 The following items are compatible with your CCT1500 System Product SKU Arm Leg Wrap Right Regular AQU637 RS1 Arm Leg Wrap Left Regular AQU638 LS1 Aquilo Recovery Pants Size 1 AQU626 S3R Aquilo Recovery Pants Size 2 AQU627 S3R Aquilo Recovery Pants Size 3 AQU628 S3R Aquilo Recovery Pants Size 4 AQU629 S3R Carry Case AQU661 DB1 Wrap Bag AQU664 WB1 Aquilo Heater US AQU635 H...

Страница 5: ...l Unit Size 410 x 210 x 260 mm Weight 8 2 kg CCT1500 Control Unit Compression Boots Pressure level 20 mmHg up to 150 mmHg AC power 100 240 V 50 60 Hz 2 5A Max DC input 13V 6 5A Battery Life 2 pumps for 1 hr 1 pump for 3 hrs ...

Страница 6: ...wer supply The battery can power both the air and water pumps for approximately 1 hour or power either pump independently for approximately 3 hours Battery Storage Store in a cool preferably below 30ºC and ventilated area away from moisture sources of heat open flames and food and drink Temperature above 90ºC may result in battery leakage and rupture Do not incinerate deform mutilate crush pierce ...

Страница 7: ...illed After the last chamber CH6 is filled all chambers will release and begin again Mode 2 Uniform Compression all air chambers are filled at the same time Mode 3 Sequential Compression Top to Bottom Mode 4 Sequential Compression Bottom to Top 1 chamber at a time releasing previous channels Mode 5 Sequential Compression Bottom to Top 2 chambers at a time Note The system will automatically default...

Страница 8: ...l Unit reservoir not the temperature experienced by the user The external temperature of the water pads will likely be slightly warmer than the water inside the Control Unit Note during a recovery session the water will increase in temperature as it is circulated and the ice in the Control Unit melts Continuously add ice through the recovery session s to maintain the target temperature of the pads...

Страница 9: ...Display Water Panel Controls function of the water pump Compression Panel Controls function of the air pump ...

Страница 10: ...Indicators ...

Страница 11: ...et Use this button to select one of the available Modes 1 5 as well as time and pressure settings Play Pause Use this button to start or pause a compression treatment Add Time Compression Use this button to add time or pressure Subtract Time Compression Use this button to reduce time or pressure Pressure Time Selection Use this button to confirm pressure and time once the desired number is added o...

Страница 12: ...sing the provided power supply The power supply air tubing and water tubing pose potential tripping hazards Position the device to mitigate these risks to yourself and others External environmental factors will impact the performance of this device High external temperatures will increase water temperatures faster than normal Cryo Compression Treatment 1 Open the reservoir door using the finger gr...

Страница 13: ... buttons Tap the OK button again to finalize selections 8 To prevent compression of sensitive areas or focus the treatment on a specific area compression chambers can be excluded from the program To exclude a chamber select that chamber on the chamber selection diagram 9 Place the user s legs into the Boots sliding each Boot up their leg until their heel is squarely at the bottom Completely zip th...

Страница 14: ...ssion of sensitive areas or focus the treatment on a specific area compression chambers can be excluded from the program To exclude a chamber select that chamber on the chamber selection diagram 5 Place the user s legs into the Boots sliding each Boot up their leg until their heel is squarely at the bottom Completely zip the Boots and ensure the user is seated comfortably prior to starting the tre...

Страница 15: ...sconnecting water tubing to let the water drain out of the pads and back into the reservoir 10 Press the power button to tun off the unit Unplug the power supply and water tubes from the control unit 11 With the unit placed securely at the edge of a sink or other drain untwist a drain cap on the side of the unit to let the water flow out of the reservoir Replace the drain cap after the water has b...

Страница 16: ...stem cleaning method The exterior of the water and air tubes can be cleaned using a mild detergent or a mixture of water and isopropyl alcohol Apply the cleaner to a cloth and wipe down the surfaces of the tubes Allow the tubing sets to dry completely before returning it to storage or the carrying bag Avoid using harsh cleaners or abrasive materials that could damage the plastics Avoid using any p...

Страница 17: ...and restart the treatment session The Control Unit appears to be leaking Water has leaked out of the inner reservoir into the insulating outer shell Drain all water out of the unit Let the tank completely dry out Make sure the drain caps are tightly fastened to the sides of the unit Try to use the device again If the problem persists contact Aquilo Sports The tubing connectors are not properly con...

Страница 18: ...g power The device battery is dead Make sure the device is plugged into an active outlet There should be a green light on the power supply to indicate the power supply is receiving power Select a different outlet if the power supply is not receiving power If the problem persists contact Aquilo Sports The compression boots aren t reaching inflation pressure Air tubing is disconnected not connected ...

Страница 19: ...Only use the included power supply to charge and power the device Switching Power Supply Model HG85 C13P Do not use any liquids other than water in the Control Unit reservoir Doing so could result in damage to the Control Unit and or water pads Use only Aquilo Sports water pads and compression boots with the CCT1500 Control Unit Only use the handles on the side of the product to carry the Control ...

Страница 20: ...perly covered with a sterile dressing prior to treatment Discontinue use if the user experiences discomfort or numbness while using the device Read and understand all caution warning labels on the device Do not remove labels from the device Indications for Use The CCT1500 System combines cold and compression therapy modalities into a single device The device allows for full or partial coverage of ...

Страница 21: ...ombosis Suspected observed inflammatory phlebitis process Suspected observed pulmonary edema Suspected observed acute inflammations of the veins thrombophlebitis Suspected observed decompensated cardiac insufficiency Suspected observed arterial dysregulation Suspected observed erysipelas Suspected observed carcinoma and carcinoma metastasis in the affected extremity Suspected observed decompensate...

Страница 22: ...tly aquilosports com The user will receive an RMA Returned Material Authorization form to complete If you return the CCT1500 Control Unit to Aquilo Sports you must assume the risk of damage or loss during shipping You must use the original packaging or the equivalent Aquilo Sports may require you to verify in writing that you are the original purchaser of the CCT1500 Control Unit Aquilo Sports may...

Страница 23: ...LE AND EXCLUSIVE REMEDIES FOR ANY BREACH OF WARRANTY AQUILO SPORTS IS NOT RESPONSIBLE FOR DIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY BREACH OF WARRANTY OR UNDER ANY OTHER LEGAL THEORY INCLUDING BUT NOT LIMITED TO LOST PROFITS DOWNTIME GOODWILL AND DAMAGE TO OR REPLACEMENT OF EQUIPMENT AND PROPERTY SOME JURISDICTIONS DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL ...

Страница 24: ...put lines 2 kV for power supply lines 1 kV for input output lines Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on pow...

Страница 25: ... to prevent electromagnetic radiation from the CCT1500 device from interfering with other equipment in the vicinity Interference may be decreased by using a separate electrical outlet increasing physical distance between devices and reorienting relocating the device WARNINGS The Aquilo Sports CCT1500 device should be used in accordance with the guidelines outlined within the Electromagnetic Compat...

Отзывы: