background image

AT-G-DETECT

Gas Sensing System

LEAK DETECTION SOLUTIONS

3

D. BUMP TEST

Ideally bump tests are conducted on site in a 
clean air atmosphere.

SEMICONDUCTOR SENSORS

We o

 er cylinders of calibration gas at known 

concentration and ampoules for ammonia 
(NH3) at 100ppm and1.000 ppm and using 
these constitute a quanti

fi

 ed bump test. 

 

the gas onto the sensor and force it into 
alarm. Check that alarm lights and relays are 
activated.

INFRARED SENSORS FOR CO2 
DETECTION

For a quanti

fi

 ed bump test you can check the 

infrared sensors for carbon dioxide using 
Aquilar ampoules 

fi

 lled with CO2 at 2000ppm 

in air or calibration cylinders. If these are not 
available, you can carry out a non-quanti

fi

 ed 

bump test by breathing on the sensor. The 
human breath has enough CO2 to trigger the 
alarm. If the bump test is not successful then 
carry out a gas calibration, see below.

E. BUMP TEST USING GAS AMPOULES

1. Make sure that both the ampoules and the

calibration beaker are clean and dry.

2. Unscrew the beaker hold screw and place

the ampoule so that it sits in the base of
the beaker

3. Tighten on the screw onto the ampoule

without breaking it

4. Remove the enclosure lid of the gas sensor

(not in Ex area and in one level units as

monitoring of voltage can be done on 
controller). If this is a CO2unit also remove 
the yellow cap (if on your model) on the 
calibration port

5. Connect volt meter to monitor sensor

response

6. Place the beaker over the sensor head

(using an adaptor if required)or, if an
Exd or Remote sensor head version, M35
or M42 thread, screw the beaker on the
remote sensor head. It should be as tight

fi

 tting as possible to allow maximum

exposure to the gas.

7. CO2 ampoule: hold the beaker in a

45-degree angle as per illustration. This
allows gas to 

fl

 ow through the front back

of the sensor and though the calibration
ports).

Unit 30, Lawson Hunt Industrial Park, 

Broadbridge Heath, Horsham, West Sussex, 

RH12 3JR
+44 (0) 1403 216100
[email protected]
www.aquilar.co.uk

Содержание AquiTron AT-G-DETECT

Страница 1: ...AT G DETECT Refrigerant Gas Sensing System Check Calibration Procedure AquiTron CA L I B R AT I O N P R O C E D U R E...

Страница 2: ...ective is to establish if the sensor is reacting to the gas and all the sensor outputs are working correctly A quantified bump test is one where gas of a know concentration is used CALIBRATION This co...

Страница 3: ...r does not go into alarm carry out a gas calibration If the factory set point information is not on the calibration label as with older units check the serial number of your gas detector on the rating...

Страница 4: ...OULES 1 Make sure that both the ampoules and the calibration beaker are clean and dry 2 Unscrew the beaker hold screw and place the ampoule so that it sits in the base of the beaker 3 Tighten on the s...

Страница 5: ...o the controller to monitor sensor response Expose the sensor to gas from the cylinder You can place the entire sensor into a plastic bag or use a plastic hose hood to direct gas to the sensor head G...

Страница 6: ...3 minute alarm delay on a one level system by moving the jumper link at JP1 to the o position Two adjustments are required and they are all performed on the controller unit A Sensor Standby Voltage S...

Страница 7: ...used drift B Alarm Threshold Voltage ATV Low Treshhold Connect voltmeter between TP5 0V TP2 Ve set the voltage asshown on the calibration label by adjusting pot P8 High Threshold Connect voltmeter bet...

Страница 8: ...r extreme humidity or temperature conditions you may adjust the settings to compensate One Level Systems You should reduce the SSV level in 0 5V increments until the condition clears Two Level Systems...

Страница 9: ...libration label You can monitor this as follows 1 level systems Connect a Voltmeter and monitor voltage between Pins 4 Ve 2 Ve of the sensor terminal connector block for each channel in turn CN1 CN2 E...

Отзывы: