Set reference and
wait for adjustment
Read Value 1390 µS
Calib. value 1410 µS
1/1
(Fig. 18)
Calibration OK
Press any key to
continue
Calibrate Sensor
This menu allows you to calibrate the Conductivity probe
through the interface. By using the keys
, you may
choose whether to perform a new calibration by selec-
ting
“New”
, or to delete a previous one and reset the
interface to the default settings by selecting
“Reset”
(Fig. 16). A calibration should be performed when the
interface is first connected to the system.
Note:
Before performing the calibration, the probe
must be rinsed using tap water, dried carefully and
inserted in the Aquatronica conductivity solution.
In order to correctly calibrate the sensor, proceed as
follows:
Main screen
Main menu
Conductivity
Calibrate
Sensor
New.
1)
Select the function
“New”
by using the keys
then press
“Enter”
(Fig 15).
2)
Using the keys
set the value of the standard
solution next to
“Calib. Value”
(Fig. 17).
3)
Wait 10 minutes in order to allow the probe’s read
value to stabilize. The read value may stabilize on a
value that is slightly different than the reference.
4)
Once 10 minutes have elapsed, press
“Enter”
.
5)
The controller will display the calibration result (Fig.
18); rinse the probe and insert in aquarium
Note:
the calibration may be cancelled at any time
by pressing
“Esc”
. This will return the calibration
parameters to those of the last completed calibration.
(Fig. 15)
Calibrate Sensor
New
Reset
(Fig. 16)
Do you want to return to
default calibration values?
Enter: Confirm
Esc: Cancel