32
VCS - Viscomaster™
7.4
Disposal
At the end of its life cycle, this product should be disposed of according to local regulations
regarding waste recycling or disposal.
Batteries and rechargeable batteries shall be recycled separately.
The separate collection and recycling of used products will help to conserve natural resources,
and ensures that they are disposed of in a way that does not cause damage to the environment
and nature.
8
Technical Data
8.1
Viscomaster
Viscosity Sensor
Hardware
Body
Aluminium alloy Red
(-40°C < Ta < +110°C)]
Pressure Range
PN16
Placement
Tines are contained in a side pocket
off the main flow, recessed by 25.4
mm (1 in)
Process connection
1 1/2-inch Cone Seat Compression fitting 316/316L
Cable duct
½” NPT
Operating Range
Viscosity
Up to 100 cSt
Density
Up to 3000 kg/m³
Temperature
–50 °C to 200 °C
Ambient temperature
-40…60°C (Electronics)
Flow rate
10 to 330 l/min
Max. Pressure
207bar
Electrical
Supply range
24VDC
Power consumption
0.65 W typical, 1.1 W maximum
Output
2x 4-20mA Signal
RS 485 Terminal
Sensor Range and
Performance
Viscosity accuracy ±0.2cP (0-10cP range) then ±1% FS of
calibrated range, Viscosity limit 20,000cP
Signal Output (dependent on sensor type)
Full
Viscosity mA1+
4mA = 0cSt - 20mA = 50cSt
Density mA1+
4mA = 500kg/m³ - 20mA = 1500kg/m³
Dynamic
Viscosity mA1+
4mA = 0cSt - 20mA = 50cSt
Temperature mA2+
4mA = 0°C - 20mA = 200°C
Density
Density mA1+
4mA = 500kg/m³ - 20mA = 1500kg/m³
Temperature mA2+
4mA = 0°C - 20mA = 200°C
Содержание Viscomaster
Страница 33: ...VCS Viscomaster 33 8 2 Dimensional drawings Detailed dimensional drawing is made in project drawings ...
Страница 35: ...VCS Viscomaster 35 ...
Страница 36: ...36 VCS Viscomaster ...
Страница 37: ...VCS Viscomaster 37 For connection with external power supply controller and sensor passive ...
Страница 38: ...38 VCS Viscomaster ...
Страница 39: ...VCS Viscomaster 39 ...
Страница 47: ...VCS Viscomaster 47 ...