MAINTENANCE
7.11
FAULT SYMPTOM
FAULT
ACTION
Samples not being kept
at 5
0
C
Vent(s) blocked
Clean Vents (See page
7.1)
Lower Compartment
Door not fully closed
Close Door (or if the
door has become
missaligned, or the
Door Seal damaged,
request On-Site
service)
Condensate Drain
blocked
Unblock Drain (See
page 7.2)
Refrigeration system
faulty
Request On-Site
service
In addition to the above Fault Symptoms the Sampler is able to
detect and register various Warning Events. (See page 6.44 for more
information). The following is a list of Warning Events with the Log
Shot Form (which appears when the Warning log is examined, by
pressing the View button), explanatory notes and a suggested course
of action
IDENT
EVENT
LOG SHORT
FORM
NOTES
ACTION
WE1
SHOT TIMED
OUT
SHT T/OUT
The Sample
Chamber
Electrodes
were not
bridged by
wastewater
after a 68
sec intake
phase.
1) Ensure the
Intake Filter is
fully immersed.
(see Intake
Hose
Installation,
page 5.14).
2) Clean the
Sample Tract.
(see page 7.2).
Sampler in
PPI Mode
Check if
the Heating
System Fault
light is on.
Cont....