OP
ERA
TION
4.11
Special Options Settings
An additional menu of Special Options is available for functions which are only required
for specialised applications. This menu can only be accessed when DEFAULT screen
To enter the SPECIAL OPTIONS menu press and hold
‘
’
button for at least 2 seconds
then release and briefly press
‘
✓
’
button
IDENT
DISPLAY
ACTION / COMMENTS
SPECOPDEF
SPECIAL OPTIONS
YES
[EXIT]
Select YES to see options available or EXIT to return to
default display
➔
See SPECOP1 if YES
➔
See DEFAULT if EXIT
SPECOP1
AUTO RESTART
NO
[YES]
The Sampler can be made to automatically restart its
program after a power interruption (this option is not
available when a Bottler is connected)
➔
See SPECOP2
SPECOP2 *
SHOTS/BOTTLE(G)
NO
[YES]
The Sampler can be made to deposit individually designated
numbers of sample shots into each bottle of a bottle array
when in GROUP mode
➔
See SPECOP3
SPECOP3
SAMP ACQUIRED OP
ON COMPLETION
[DURING ATTEMPT]
‘
Sample Acqui
red’ relay output can be programmed to
operate in 2 ways:
•
ON COMPLETION
–
Closes on acquisition of a
successful sample shot, at the end of the INTAKE phase
(SPOTSAM3)
•
DURING ATTEMPT
–
Closes for the full duration of the
sample shot attempt (whether successful or not),
opening at the end of the POST PURGE phase
(
SPOTSAM4)
➔
See SPECOP4
SPECOP4
SAMP WARNING OP
NO LATCH
[AUTO OFF]
[MANUAL]
‘
Sampler Warning
’ relay output can be programmed
to
operate in 3 ways:
•
NO LATCH - Do not latch, the output will hold for 30
seconds only
•
AUTO OFF - Latch and un-latch automatically after a
pre-set time, or when INFORMATION button is
pressed
•
MANUAL - Latch and only un-latch when
INFORMATION button is pressed
➔
See SPECOP5 if AUTO OFF
Otherwise:
➔
See SPECOPDEF
SPECOP5
AUTO LATCH TIME
00HR 00MIN 30SEC
Set the
‘
Sample Warning
’ relay output
latch time
➔
See SPECOPDEF
Содержание Aquacell P2 Series
Страница 1: ...AQUACELL PORTABLE WASTEWATER SAMPLERS USER GUIDE Sira MC 050059 Issue 20 2...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ...CONTENTS OF PACKAGE 1 2...
Страница 14: ...CONTENTS OF PACKAGE 1 8...
Страница 16: ...INTRODUCTION 2 2...
Страница 26: ...INSTALLATION 3 2...
Страница 36: ...INSTALLATION 3 12...
Страница 38: ...OPERATION 4 2...
Страница 56: ...OPERATION 4 20...
Страница 58: ...MAINTENANCE 5 2...
Страница 70: ...TECHNICAL SPECIFICATION 6 2...
Страница 77: ......