OP
ERA
TION
4.6
SYSSET4
SAMPLE TEMP LOG
NO
[STOP WHEN FULL]
[ROLLING]
The Sampler is capable of logging sample temperatures
when specified with Sample Temperature Monitoring (see
MAINTENANCE for details of fault registration). There are
2 ways in which the Sample temperature data can be
stored:
•
ROLLING
–
Where data will over write when the log is
full, giving approximately the last 41 days of data for
up to 70 days from running the program
•
STOP WHEN FULL
–
Where data will stop being
stored when the log is full, giving the first 30 days of
data only
➔
See SYSSET5
SYSSET5
LOW POWER MODE
NO
[YES]
When selected, in the absence of a button press (for
approx. 70 seconds), the Sampler will enter a low power
consumption mode. During this phase the LCD will switch
off. Pushing any button will restore the LCD display.
Notes:
•
LOW POWER MODE cannot be used when
SAMPLING MODE is set to FLOW
•
LOW POWER MODE will disable the Sampler Warning
Output
➔
See SYSSET6
SYSSET6
PPI MODE
NO
[YES]
The Sampler is capable of sampling from a pressurised
effluent source when specified with the appropriate
Pressurised Pipeline Interface / Connection
➔
See SYSSAM1
SYSSAM1
SAMPLER SETTINGS
> > > > > > >
Information screen only
➔
See SYSSAM2
SYSSAM2
BOTTLER FITTED
NO
[YES]
The Sampler is capable of distributing samples into
multiple bottles when specified with a Bottler
➔
See SYSSAM3
SYSSAM3
SHOTS/SAMPLE
01
The Sampler is capable of extracting larger volumes of
effluent by taking multiple consecutive sample shots each
time a Sample is due
➔
See SYSSAM4
SYSSAM4
INITIAL SAMPLE
YES
[NO]
You can choose to take a sample either at the beginning
of a routine or after the first time interval / flow increment.
When the application is to obtain flow proportional
samples then the samples should always be taken after a
certain volume of water has passed the sampling point
and therefore an initial sample should not be taken. Where
the requirement is to take samples from a sampling point
at equal time intervals commencing at a certain time then
an initial sample is sometimes appropriate
➔
See SYSSAM5
Содержание Aquacell P2 Series
Страница 1: ...AQUACELL PORTABLE WASTEWATER SAMPLERS USER GUIDE Sira MC 050059 Issue 20 2...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ...CONTENTS OF PACKAGE 1 2...
Страница 14: ...CONTENTS OF PACKAGE 1 8...
Страница 16: ...INTRODUCTION 2 2...
Страница 26: ...INSTALLATION 3 2...
Страница 36: ...INSTALLATION 3 12...
Страница 38: ...OPERATION 4 2...
Страница 56: ...OPERATION 4 20...
Страница 58: ...MAINTENANCE 5 2...
Страница 70: ...TECHNICAL SPECIFICATION 6 2...
Страница 77: ......