Technical Data
Technical Data
Instrument
three
wavelength, automatic wavelength selection,
direct reading colorimeter
Light source:
LEDs,
interference filters (IF) and photosensor in transparent
cell chamber. Wavelength specifications of the IF:
430 nm
= 5 nm
530 nm
= 5 nm
560 nm
= 5 nm
Wavelength accuracy
± 1 nm
Photometric accuracy*
3% FS (T = 20° C – 25° C)
Photometric resolution
0.01 A
Power supply
4 batteries (AAA/LR 03)
Operating time
17hr operating time or 5000 measurements in constant
use when display backlight and Bluetooth
®
data transfer
are off.
Auto-OFF
automatic switch off
20 minutes after last keypress
Display
backlit LCD (on keypress)
Storage
internal ring memory for 125 data sets
Serial Interface
Bluetooth
®
for the transfer of stored measurement data
Bluetooth
®
Modul:
Specifications:
Modul:
BLE113-A
Bluetooth
®
4.0 LE
FCC ID: QOQBT113
IC:
5123A-BGTBLE113
Time
real time clock und date
Calibration
user and factory calibration
resetting to factory calibration possible
Dimensions
155 x 75 x 35 mm (LxWxH)
Weight
approx. 260 g (incl. batteries)
Ambient conditions
temperature: 5– 40°C
rel. humidity: 30–90% (non-condensing)
Waterproof
floating; as defined in IP 68 (1 hour at 0.1 meter)
CE
Certificate for Declaration of CE-Conformity
at www.aqualytic.de
*measured with standard solutions
To ensure maximum accuracy of test results, always use the reagent systems supplied by
the instrument manufacturer.
Hı
Lo
btLo
Operating messages
Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Replace batteries, no further tests possible.
Battery capacity is too low for the display backlight;
measurement is still possible.