45
AL110_2 03/2018
GB
Methods
Notes
:
1. Before use, clean the vials and the accessories with Hydrochloric acid (approx. 20%). Rinse
them thoroughly with deionised water.
2. To get accurate results the sample temperature must be between 20°C and 25°C.
3. A low test result may be given in the presence of Fluorides and Polyphosphates. The ef-
fect of this is generally insignificant unless the water has fluoride added artificially.
In this case, the following table should be used:
Fluoride
[mg/l
F]
Displayed value: Aluminium [mg/l Al]
0.05
0.10
0.15
0.20
0.25
0.30
0.2
0.05
0.11
0.16
0.21
0.27
0.32
0.4
0.06
0.11
0.17
0.23
0.28
0.34
0.6
0.06
0.12
0.18
0.24
0.30
0.37
0.8
0.06
0.13
0.20
0.26
0.32
0.40
1.0
0.07
0.13
0.21
0.28
0.36
0.45
1.5
0.09
0.20
0.29
0.37
0.48
---
Example: If the result of Aluminium determination is 0.15 mg/l Al and the Fluoride con-
centration is known to be 0.4 mg/l F, the true concentration of Aluminium is 0.17 mg/l Al.
Reagent
Form of reagent/Quantity
Order-No.
Set
VARIO Aluminium ECR F20
VARIO Aluminium Hexamine F 20
VARIO Aluminium ECR Masking Reagent
Powder Pack / 100
Powder Pack / 100
Liquid reagent / 25 ml
4535000