quality of resin we use, the water you get will be, literally as pure as possible.  The 

resin is a color indicating resin; it will slowly change from a dark purple/blue/black 

to a rust red/orange color.  As this filter nears the end of its useful life you may 

experience “break though”.  Keep this in mind, depending on the water quality 

needed. For applications demanding 18 Mega-Ohm water, 2 DI’s are STRONGLY 

recommended. 

 

 

Trouble shooting guide 

  

Problem

 

Solution

 

High TDS after 

membrane 

 

Clogged Pre-filters. Causing Pressure Drop. Insufficient pressure will yield 
poor TDS rejection from RO membranes. Change Pre-Filters.  
Average Membrane life is 2 to 3 years. Most common cause for 
membrane failure is insufficient pre-filter maintenance. AquaFX Carbons 
are rated for 3,750 gal @ 1.0 GPM of 2ppm Chlorine. Do not exclude 
drain water from this capacity, as it is treated water. Very high TDS 
(>550ppm) may also yield premature failure.

 

Low incoming water 

pressure.

 

Incoming water pressure must be above 40 PSI, Add a booster pump if 
below 40psig. Make sure Pre-Filters are not clogged causing low 
pressure.

 

No water to drain.  

 

Flow Restrictor is clogged, Replace Drain Flow Restrictor. (rare) 
Check to see if water is turned on.

 

All water is going 

out of the drain  

 

With no back pressure, most of the water will exit out of the drain. Make 
sure a Flow restrictor is present. 

 

Very little/slow 

water production  

 

Reference units flow rate vs. actual production. Slow flow from RO is 
normal. Cold water, low pressure and high contaminant levels will 
contribute to decreasing the rate of water production. Heating water (NO 
WARMER THEN 80 Degrees F), increasing pressure or additional pre-
filtration will help to counteract these adverse affects.

 

Drain Water 

Continues After 
Product line is full

 

The most common cause of a ‘continuing drain’ is a pressure leak on the 
product side of the RO. If the pressure cannot build, the automatic shut-off 
(ASO) will not close. Check ASO with a ball valve right after RO. Close 
Ball valve then wait 1-3 minutes. If drain continues, replace ASO Valve. If 
it stops, have the customer, find then stop pressure leak. If a float valve is 
being used, it is more than likely the culprit.

 

The incoming feed 

water TDS has 
increased.

 

An increase in feed water TDS will also give an increase in Product Water 
TDS.  R.O. (Dolphin) Users will see this rise in TDS, due to the membrane 
removes a percentage of TDS. 

 

 

  DI is 

exhausting 
very fast 

R.O. Membranes are manufactured with a preservative on them, if your initial 
startup allows the membrane to ‘rinse’ into the DI; there will be some immediate 
exhaustion, followed by normal exhaustion. 

 
Chloramines Vs Chlorine 

– If the customer’s water is disinfected with 

Chloramines (NH

2

CL) the ammonia portion will still be present post RO. 

 

Содержание 3 Stage Dolphin RO

Страница 1: ...olphin RO Instruction Owner s Manual 7208 Aloma Ave Suite 100 Winter Park FL 32792 Phone 407 599 2123 Fax 407 599 2124 www Aquariumwaterfilters com www AquaFX net 7208 Aloma Ave Suite 100 Winter Park...

Страница 2: ...ft side of the system 6 Once installed allow the system to run for approximately 15 minutes to flush membrane completely Please check systems for leaks parts can come lose during shipping 7 You will k...

Страница 3: ...The Effect of Temperature on Permeate Production using a 50 GDP TFC Membrane...

Страница 4: ...water temperature 50 GPD is produced at approximately 72 degrees F at a constant pressure of 50 PSIG and constant TDS of 125PPM Sodium softened water Temp F GPD Factor 77 50 1 68 42 85 875 59 36 5 730...

Страница 5: ...s which require custom materials or components i e custom tanks brackets electrical components cannot be cancelled once special material has been ordered and or production has begun Accessories Drinki...

Страница 6: ...supplies ONLY Warranty begins at the time of product registration and must be registered within 10 days of the date of purchase This warranty does not require replacement of the entire unit The defect...

Страница 7: ...pm Chlorine Cl2 The Chloramine Blasters are run 2 in series and will yield approximately 8000 gallons of treated water Do not exclude drain water from this capacity as it is treated water 3 Reverse Os...

Страница 8: ...he water will exit out of the drain Make sure a Flow restrictor is present Very little slow water production Reference units flow rate vs actual production Slow flow from RO is normal Cold water low p...

Страница 9: ...se DI to remove the ammonia RO DI DOLPHIN Mako and Great White users will have the ammonia removed by the DI This will tax the DI Chloramines appears in several forms so the decreased life would depen...

Отзывы: