Preparations Before Treatment
1. It is recommended that a light coating of petroleum jelly is smeared around the
patient’s lips to prevent cracking and to retain some of the outflowing abrasive
powder.
2. Rubber dam should be used to protect adjacent teeth and soft tissue.
3. Non-metal Matrix strips should be used to protect adjacent teeth during interproximal
preparations.
4. A bib should be used to protect the patient’s clothes from any out-flow of the
material from the mouth during the course of a treatment.
5. Use a high-speed intra-oral aspirator, preferably with a flared tip.
6. An extra-oral evacuator can also be used with the extraction cone sited below the
patient’s chin.
7. IMPORTANT: To avoid unnecessary damage, when switching from Aluminium
Oxide to Sodium Bicarbonate you must purge the powder. See MAINTENANCE:
PURGING THE AQUACARE SINGLE, Pg 33.
8. It is recommended that AquaCare Cleaning/Cutting Fluid (AquaSol) should be used
in the AquaCare Fluid System. The 500ml bottle supplied with the machine should
be discarded when empty and replaced. The fluid can be used with all recommended
treatments using this instrument.
AquaCare: Operation
Please read INTRODUCTION: SAFETY WARNINGS, Pg 5-6 carefully before using
the AquaCare.
WARNING: All people in the operating area must wear eye protection to BS EN
166 IF 415
WARNING: All people in the operating area except for the patient must wear
respiratory masks to EN 149 FFP2S