1HP & 2HP Fusion Series Owner’s Manual
Page 7
Fusion Series Troubleshooting G uide
TROUBLE
CHECK
POSSIBLE SOLUTIONS
Fusion does not operate
-OR-
Starts then stops
Not plugged in
Cord must be plugged into a 3-wire outlet
that is protected by a GFCI (Ground Fault
Circuit Interrupter).
GFCI/RCD Tripped
Reset GFCI/RCD. If GFCI/RCD will not
reset or trips immediately, unplug the pump
& lights and reset GFCI/RCD. If GFCI/RCD
will still not reset, there may be a problem
with the incoming power. If the GFCI/RCD
resets, plug the pump. If the GFCI/RCD
trips, there may be a problem with the pump
cord. If the GFCI/RCD holds, plug in the
optional light kit. If the GFCI/RCD trips,
there maybe a problem with the lights.
Contact Aquatic Technologies Technical
Support totroubleshoot lights or pump
.
GFCI/RCD not
tripped,not
running.
Disconnect power. Check propeller for
obstruction. These motors have internal
overloads. If the propeller is jammed, the
internal overload may trip. This overload
may require up to 10 minutes to cool down
& reset. If the motor does not restart, there
may be other motor problems. Contact
Aquatic Technologies Technical Support to
troubleshoot the pump.
Fusion spray pattern is not
reaching full size
Fusion intake clogged Disconnect power. Check suction screen for
obstruction & remove debris.
Nozzle clogged
Disconnect power. Remove nozzle &
check for obstruction & remove debris.
A quatic Technologies
41 Yazaki Way, Carrum Downs 3201
Ph: (03) 9071 2442
aquaticenquiries.aquatic.net.au
-OR-
Contact Your Local Distributor