Document 4207Q258 V3.00 January 2018
Page 7 of 30
Chirascan Couette Cell LD Accessory User Manual
CHIRASCAN COUETTE CELL LD ACCESSORY INSTALLATION AND OPERATIONAL
REQUIREMENTS
Environmental and electrical requirements
The Couette Cell LD Accessory has no environmental requirements additional to those of the Chirascan
spectrometer. The accessory has a power unit that requires a separate mains electricity supply.
Bench space
The Couette Cell LD Accessory power supply has a footprint of about 260 x 260 mm and should stand to the
right of the spectrometer sample chamber.
Nitrogen purge gas
The Couette Cell LD Accessory has no purge gas requirements additional to those of the Chirascan
spectrometer.
Circulating water
For temperature control, the Couette Cell LD Accessory requires a circulating water flow rate of about 200
–
300 ml per minute. To achieve this, a pressure of approximately 0.2
– 0.3 bar (3 to 4.5 p.s.i.) is needed. For
further information on suitable circulators, please contact the Applied Photophysics Technical Support
Department at the address given on the front of this User Manual.
Servicing
Servicing of the Couette Cell LD Accessory should only be undertaken by qualified personnel. If you are in any
doubt at all please contact the Applied Photophysics Technical Support Department at the address given on
the first page of this User Manual.