Example: Calf thymus DNA
Linear dichroism spectra for calf thymus DNA, Sigma-Aldrich product D1501, at shear rates from 0 to 10,200
s
-1
, are shown in Figure
. The spectrum for a shear rate of 0 s
-1
was used as the baseline. The conditions
used are shown in Table
Figure 3.13: LD spectra for calf thymus DNA at various shear rates
Sample
Calf Thymus DNA
Concentration
0.13 mg/ml
Buffer
Water
Wavelength range
180 to 360 nm
Wavelength step
1 nm
Bandwidth
1 nm
Time-per-point
0.5 s except shear rates 120 and 240 s which were 1 s
Pathlength
0.5 mm (total)
Temperature
Ambient (
20.5
C)
Repeats
3 (shear rates 0 to 30 s
-1
)
2 (shear rates 60 to 10,200 s
-1
)
Shear rates
0 (baseline), 1, 2.4, 6, 12, 30, 60, 120, 240, 480, 960, 1920, 3840,
6000, 7800, 9000, 10,200 s
-1
Table 3.1: Conditions used for the LD spectra of DNA at various shear rates