Note:
Confirm that the replacement heated cover supports the plate or array card
format that you want to use. Some heated covers support more than one consumable
type.
1.
Power off and unplug the QuantStudio
™
6 or 7 Flex Real-Time PCR System, then
allow it to cool for 15 minutes.
2.
Open the access door.
3.
Unlock the heated cover by pinching the handle together, then pull the assembly
from the system and place it on a clean, dry surface.
4.
Install the new heated cover into the QuantStudio
™
6 or 7 Flex Real-Time PCR
System.
IMPORTANT!
Align the arrows on the front handle as shown below. If the
arrows are not aligned, push the heated cover into the system until the handle
locks into place.
Correct
Incorrect
5.
If you are installing a heated cover of a different consumable format, replace the
sample block and plate adapter if necessary.
Change the heated
cover
Appendix B
Maintenance
Change the heated cover