AT525D User's Guide
Example
SEND> FUNC:RATE?
<NL>
RET>
SLOW
<NL>
8.8
COMParator Subsystem
The COMParator
subsystem command group sets the comparator function,
including its ON/OFF setting, limit mode, and limit values.
Figure 8-4 COMParator Subsystem Command Tree
COMParator
:BEEP
{OFF,GD,NG}
:RMODe
{OFF,SEQ,PER,ABS}
:VMODe
{OFF,SEQ,PER,ABS}
:TOLerance
RNOMinal
<float>
VNOMinal
<float>
RLIMIT(RLMT)
<LOWER,UPPER>
VLIMIT(VLMT)
<LOWER,UPPER>
8.8.1
COMParator:RMODe
The :COMParator:RMODe command sets resistance limit mode of the
comparator function.
Command
Syntax
COMParator:RMODe {OFF,ABS,PER,SEQ}
Parameter
Where,{OFF,ABS,PER,SEQ} is:
ABS
Absolute tolerance mode
PER
Percent tolerance mode
SEQ
Sequential mode
Example
SEND> COMP:RMODe PER
<NL>
Query
Syntax
COMParator:RMODe?
Query
Response
{off,abs,per,seq}
Example
SEND> COMP:RMODe?
<NL>
RET>
abs
<NL>
8.8.2
COMParator:VMODe
The :COMParator:VMODe command sets voltage limit mode of the comparator
function.
Command
Syntax
COMParator:VMODe {OFF,ABS,PER,SEQ}
Parameter
Where,{OFF,ABS,PER,SEQ} is:
ABS
Absolute tolerance mode
PER
Percent tolerance mode
SEQ
Sequential mode
Example
SEND> COMP:VMODe PER
<NL>
Query
Syntax
COMParator:VMODe?
Query
Response
{off,abs,per,seq}