![Applent Instruments AT525D Скачать руководство пользователя страница 25](http://html.mh-extra.com/html/applent-instruments/at525d/at525d_user-manual_2954886025.webp)
<SYSTEM CONFIG> page
25
6.1.6
RS-232 Result Send Mode [RESULT SEND]
SCPI Command: SYSTem:SENDmode {FETCH,AUTO}
When you set the [RESULT SEND] to AUTO, the test result will be sent to host
every end of measurement instead of by sending “FETCH?” command.
The format is:
+3.549568e-01,+3.00,RV GD
+3.549911e-01,+3.00,RV GD
+1.20,+1.20,RV NG
…
Where,
“+1.20” stands for overload or open.
“RV GD” stands for Resistance and Voltage are Good (GD).
“RV NG” stands for Resistance and Voltage are Not Good (NG) or invalid
◼
Procedure of setup the result send mode:
Step 1.
Press the [Meas] or [Setup] key
Step 2.
Press the [SYSTEM] bottom soft-key.
Step 3.
Use the cursor key to select [RESULT SEND] field
Step 4.
Use the soft keys to turn ON.
FETCH
Acquire the test result by sending
“FETCH?” command only.
AUTO
AT525D return the result every EOM
6.2
<SYSTEM INFORMATION>
There are no configurable options in the <SYSTEM INFO> page.
Figure 6-2 <SYSTEM INFORMATION>page
◼
Procedure of viewing system information
:
Step 1
Press [Meas] or [Setup] key to enter main page
Step 2
Press bottom soft key [SYSTEM] to enter <SYSTEM CONFIG>
Step 3
Press soft-key [SYSTEM INFORMATION] to enter <
SYSTEM INFORMATION
>