16
AT525D User's Guide
Step 1
Press [Meas] key to enter measurement page
Step 2
Use the cursor key to select [RANGE]
Step 3
AUTO
Auto range
HOLD
Current range is hold
NOMINAL Select the range according to resistance nominal
value [RNom].
INCR+
Increase range
DECL-
Decline range
4.1.3
Measurement Speed [SPEED]
SCPI Command: FUNCtion:RATE {SLOW,MED,FAST}
SCPI Query Command:
FUNCtion:RATE?
SLOW, MED, FAST and ULTRA speed mode can be selected for AT525D.
SLOW mode will result in more stable and accurate measurement result.
The following speed is measured in range-hold mode and comparator is ON:
SLOW:
1 times/sec (1s)
MED:
5 times/sec (100ms)
FAST:
`0 times/sec (33ms)
•
Procedure to set test speed
:
Step 1
Press [Meas] key to enter measurement page;
Step 2
Use the cursor key to select [SPEED] field
Step 3
SLOW
MED
FAST
4.1.4
[BEEP] Feature
SCPI Command: COMParator:BEEP {OFF,GD,NG}
SCPI Query Command:
COMParator:BEEP
◼
Procedure to set the beep:
Step 1
Press [Meas] key to enter measurement page;
Step 2
Use the cursor key to select [BEEP]
Step 3
OFF
Turn Beep Off
GD
Beep while Pass
NG
Beep while Fail
4.2
Status Bar on <MEAS DISPLAY>
⚫
HOLD – Hold the data under internal trigger mode, test stops.
⚫
SYSTEM– Go to <SYSTEM CONFIG>Page
⚫
KEY LOCK – Lock the keypad.
⚫
Time