![Applent Instruments AT520 Скачать руководство пользователя страница 36](http://html.mh-extra.com/html/applent-instruments/at520/at520_user-manual_2954882036.webp)
AT520/AT520SE/AT520L/AT520M User’s Guide
36
8.10
ERRor
Subsystem
The ERRor subsystem sets or retrieves last error information.
Figure 8-6
ERRor Subsystem Tree
Query Syntax
ERRor?
Query Response
no error.
For example
:
Tx>
ERR?
Rx>
no error
Constraints
None
8.11
*IDN?
Common
Command
The *IDN? query returns AT520x Version.
Figure 8-7
IDN? Common Command Tree
Query Syntax
*IDN?
Query Response
<model>,<version>
For example
:
Tx>
*IDN?
Rx>
AT520,V2.00
8.12
*RST
Common
Command
*RST restarts instrument.
Figure 8-8
*RST Common Command Tree
Command Syntax
*RST
Parameters
none
For example
:
Tx>
*RST
Rx>
Wait for 3s…
Query Syntax
none
Constraints
none
8.13
*TRG
Common
Command
The *
TRG
common command is used to enable a measurement and fetch the test value.
*TRG = TRIG;:FETC
Figure 8-9
*TRG Subsystem Tree
RST?
IDN?
ERRor?