9
* One press on AUTO H key (LOG RATE) enables this function, the 1st
default is on sub-display, use "
▼
" or "
▲
" to select Log Rate period for
data log-in.
* The default (1st) Log Rate is adapted to measuring requirements.
* Again press this key once in this mode confirms the Log Rate for log-in
and escaps.
4.12.2 DATA LOG-IN
* Pressing on PEAK H key
≧
2 sec enables data log-in function.
* Meter starts storing measured reading referring to log rate selected into
memory.
* One press on this key in this mode interrupts data log-in with “PAUS” on
sub-display for log rate re-setting.
* While data log-in performing, the “—” sub-display is blinking for storing
indication.
* When stores data at 40Kth (6Kth for 107N), data log-in stops, the bar is
full (for 109 only) , sub-display keeps blinking “—” and showing FULL.
4.12.3 DATA LOG-OUT
* One press on PEAK H key (DATA LOG) enables data log-out function
when is just in data log mode, sub-display appears sequence number
of this logged-in reading has displayed on main display.
* The first data displayed when just enabled data log-out is the last reading
logged-in.
* In this mode, one press on "
▼
" or "
▲
" key sequentely steps up or down
the logged-in data.
* Pressing on "
▼
" or "
▲
" key
≧
2 sec performs sequentely steps up or
down the logged-in data with 10 data/sec display rate and stops at the
reading when the key is released.
* In this mode, one press M/ M/ A key toggles maximum and minimum
values logged-in, pressing on
≧
2 sec escaps.
* In this mode, press RANGE key and then press on "
▼
" or "
▲
" key
toggles the turning points logged-in with “MAX” or “MIN” symbol depending
on the comparison from current reading and next reading. Press Range
key again to escape.
4.12.4
▲
: Scrolls up data of function selected.
4.12.5
▼
: Scrolls down data of function selected.