background image

11 

5 OPERATION 

5.1 VOLTAGE MEASUREMENTS (AC, DC, AC + DC) 

(Set to autoranging mode  

      for unknown voltage measurements). 

     * Set dial. 

      * Choose AC, DC or AC + DC

 

    * Connect leads 

* The AC and AC + DC measurements provide a true RMS measurement. 

* In AC mode, the frequency of the measured signal 

20% of full scale is  

   displayed on sub-display.  

Содержание 107N

Страница 1: ...32 APPA 107N 109N DIGITAL MULTIMETER INSTRUCTION MANUAL ...

Страница 2: ... CAPACITANCE MEASUREMENTS 17 9 MAINTENANCE 28 10 ACCESSORIES 30 4 12 2 DATA LOG IN 9 5 5 CONTINUITY CHECK 15 5 6 CURRENT MEASUREMENTS AC DC AC DC 16 4 12 3 DATA LOG OUT 9 4 12 4 9 4 12 5 9 5 OPERATION 11 5 1 VOLTAGE MEASUREMENTS 11 5 2 Milli VOLTAGE MEASUREMENTS AC DC AC DC 12 5 3 OHM AND LOW VOLTAGE OHM MEASUREMENTS 13 5 4 DIODE TEST 14 5 8 FREQUENCY AND DUTY FACTOR MEASUREMENTS 18 5 9 TEMPERATUR...

Страница 3: ...ouble Insulated High Voltage Specific precautions Use proper Fuse To avoid fire hazard use only the fuse type and rating specified for this product Do not operate without covers To avoid personal injury do not apply any voltage or current to the product without the covers in place Electric overload Never apply a voltage to a connector on the product that is outside the range specified for that con...

Страница 4: ...ch type of meter has own features described in this manual The following list provides a comparison of the features between meters FUNCTION 107N DC Voltage AC Voltage mV Voltage Resistance Lo Ohm Diode Test Continuity Check DC Current AC Current Capacitance Frequency Duty Factor Temperature K Type 109N ...

Страница 5: ...r dBm Delta mode Hazard Warning Storage and Recell up to 1000 Memories Low Battery Indicator MAX MIN AVG Peak Hold 0 5ms Safety IEC UL CSA IEC UL CSA CE Mark True RMS AC DC AC VAC Hz Dual Display Water Dust Resistant Battery 9V 9V 600V High Energy Fuse LCD Backlight Auto Backlight Off 15 Minutes RS 232 Phototronic Serial Port RS 232 Cable Option Win DMM 100 software Option Holster and Stand Data L...

Страница 6: ...VIEW Figure 1 1 Input connectors 2 Measurement function dial Black labels are the initial settings blue labels are selected by the blue button 3 Function buttons 4 Data log features 5 LCD display with dual numeric readout ...

Страница 7: ...esistance indicator 5 Auto hold indicator 6 Peak hold indicator 7 Maximum Minimum and Average indicators 8 Main display unit indicators 9 Sub display unit indicators 10 STORE and RECALL indicator 11 High voltage input warning 60V DC 30V ACrms 12 is indicator low battery 13 Average indicator equals to MAX MIN 2 14 DATA LOG indicator ...

Страница 8: ... FUNCTION Prediction Buzzer beeps once for every key press Buzzer beeps twice for every invalid key press Figure 3 4 1 LIGHT This button is used to turn on or turn off the backlight This button is disabled in DATA LOG functions 4 2 BAR This button toggles analog bar center zero bar displays Zero at center Zero at center graph zoomed 2 This button is disabled in DATA LOG and non DC volt and current...

Страница 9: ...4 5 BLUE The blue button toggles between dual functions black or blue located on the dial 4 6 M M A This button toggles the MAX MIN AVG When this button is first time pressed MAX indicator is displayed and the value displayed on sub display is the most recent maximum value The beeper sounds with each new update Pressing on this key for 2 sec escapes 4 7 REL This function provides subtraction readi...

Страница 10: ...e yellow and BAR keys to scroll and veiw up and down the stored readings arrow keys perform scroll rate of 10 data sec when is pressed 2 sec and hold 4 12 RED Data Log Predictions A Data quantities 40K readings form as sequence number on sub display up to 9999 and each 1 4 scale of bar indicates 10K for 109N only B RANGE function is only enable in just data log mode C Any position measuring functi...

Страница 11: ...showing FULL 4 12 3 DATA LOG OUT One press on PEAK H key DATA LOG enables data log out function when is just in data log mode sub display appears sequence number of this logged in reading has displayed on main display The first data displayed when just enabled data log out is the last reading logged in In this mode one press on or key sequentely steps up or down the logged in data Pressing on or k...

Страница 12: ...apes Press under this mode toggles 20000 4000 count modes Rotate to power OFF then on to reset to 20000 count mode default This key is not available on following items 4 13 1 200MΩ and 2GΩ ranges of Resistances function 4 13 2 position 4 13 3 position 4 13 4 Hz DF position 4 13 5 C F position ...

Страница 13: ...et to autoranging mode for unknown voltage measurements Set dial Choose AC DC or AC DC Connect leads The AC and AC DC measurements provide a true RMS measurement In AC mode the frequency of the measured signal 20 of full scale is displayed on sub display ...

Страница 14: ...MEASUREMENTS AC DC AC DC Set dial Choose AC DC or AC DC The AC and AC DC measurements provide a true RMS measurement In AC mode the frequency of the measured signal 20 of full scale is displayed on sub display Connect leads ...

Страница 15: ...or LV Ω Connect leads CAUTION Remove all power from the circuit before connecting the test leads LV setting reduces the maximum test voltage level to about 0 5V to avoid turning on semiconductor devices Remove individual components from circuitry for best results ...

Страница 16: ...N Remove all power from the circuit before connecting the test leads Remove individual components from circuitry for best results Choose diode test Connect leads Forward bias Good 0 4 to 0 9V Bad o or 2 0V Reverse bias Good OL Bad 2 0V ...

Страница 17: ...5 5 CONTINUITY CHECK Set dial CAUTION Remove all power from the circuit before connecting the test leads The beeper sounds if the resistance of the circuit is less than 50Ω Choose diode test Connect leads ...

Страница 18: ...16 5 6 CURRENT MEASUREMENTS AC DC AC DC Set dial Choose AC DC or AC DC Connect leads ...

Страница 19: ...provide true RMS In AC mode the frequency of the measured signal 20 of full scale is displayed simultaneously 5 7 CAPACITANCE MEASUREMENTS Set dial Connect leads zero stray capacitance for low capacitance measurements CAUTION Remove all power from the circuit and discharge Capacitor to be measured before connecting the test leads Remove individual components from circuit for best results ...

Страница 20: ...18 5 8 FREQUENCY AND DUTY FACTOR MEASUREMENTS Set dial The duty factor displays the percent of the signal that is high Choose period or duty factor on sub display Connect leads ...

Страница 21: ...REMENTS Set dial This setting requires an optional temperature probe and adapter Refer to Accessories Allow about 5 minutes room temperature balance after settled for best measurement Choose Celsius or Fahrenheit Connect leads ...

Страница 22: ...urrent input connector and a measurement other than current is selected 6 3 Buzzer A single beep indicates correct operation two beeps indicate a warning or error condition 7 POWER UP OPTIONS Press button while turning meter on from OFF position LIGHT Disable backlit auto off RANGE Turn off beeper BLUE Disable Auto Power OFF AUTO H Set temperature measurement default at F M M A Clear memory area f...

Страница 23: ...layed Low voltage indicator is indicated Temperature Coefficient 0 1 x Spec Accuracy per C 18 C or 28 C Automatic power off time Automatic backslit off 15minutes Power source One 9V dry cell battery Maximum input voltage 1000V 750V AC CAT Ⅱbetween V and COM Maximum floating voltage 1000V 750V AC CAT Ⅱbetween any terminal and earth ground Maximum open circuit Voltage current inputs 600V between A a...

Страница 24: ... 80 0 70 80 1 00 80 2V 40Hz 100Hz 100Hz 1KHz 1KHz 10KHz 10KHz 20KHz 20KHz 50KHz 50KHz 100KHz 0 70 50 1 00 50 2 00 60 3 00 70 5 00 80 10 00 100 0 70 50 1 00 50 2 00 60 3 00 70 5 00 80 10 00 100 20V 40Hz 100Hz 100Hz 1KHz 1KHz 10KHz 10KHz 20KHz 20KHz 50KHz 50KHz 100KHz 0 70 50 1 00 50 2 00 60 3 00 70 5 00 80 10 00 100 0 70 50 1 00 50 2 00 60 3 00 70 5 00 80 10 00 100 200V 40Hz 100Hz 100Hz 1KHz 1KHz 1...

Страница 25: ...for C F 1 4 to 3 3 0 addition error for C F from 3 to 4 2 CURRENT DCA 107N 109N 20mA 200mA 0 20 40 0 20 40 2A 10A 0 20 40 0 20 40 ACA 107N 109N 20mA 200mA 2A 10A 40Hz 500Hz 0 80 50 0 80 50 20mA 200mA 2A 10A 500Hz 1KHz 1 20 80 1 20 80 200mA 10A 1KHz 3KHz 2 00 80 2 00 80 Range 20mA 200mA 2A 10A Resolution 1µA in the 20mA range Burden Voltage 800mV max for mA input 1V max for A input AC Conversion Ty...

Страница 26: ...0 8 LV OHM 107N 109N 2KΩ 20KΩ 200KΩ 0 60 20 0 60 30 2MΩ 0 60 50 0 60 50 20MΩ 7 00 50 7 00 50 200MΩ 7 00 20 7 00 20 excludes DIGIT switch Resolution 0 01Ω in the 200Ω range Open Circuit Voltage 3 3V Open Circuit Low Voltage 0 6V Input Protection 600V rms 5 CONTINUITY CHECK Excludes DIGIT switch Continuity Threshold Approx 50Ω Continuity Indicator 2KHz Tone Buzzer Input Protection 600V rms 6 DIODE T...

Страница 27: ...z in the 200Hz range Accuracy 0 01 10 0 01 50 for 20Hz Range Sensitivity 0 5Vp p for 5Hz 1MHz Min Frequency 5Hz Input Protection 600V rms 9 DUTY FACTOR Excludes DIGIT switch Range 20 80 Resolution 0 1 Accuracy 1 20Hz 10KHz 5Vp p 2 for 50 80 10 Temperature Excludes DIGIT switch Temperature 107N 109N 100 C 1200 C 0 1 3 C 0 1 3 C 200 C 100 C 0 1 6 C 0 1 6 C Note The readings 360 C on 1 C resolution r...

Страница 28: ...L T 28800E TYPE II Class 5 2 66gRMS 5 to 500 Hz 3axes 10 minutes each Dust Water Protection IP Rating IP 64 Indoor Use 8 5 Certifications and compliances Safety Designed to IEC 1010 1 UL3111 1 and CSA specifications Input rating 1000V DC Category Ⅱ 600V DC Category Ⅲ 750V AC Category Ⅱ 600V AC Category Ⅲ CAT Ⅲ Distribution level mains fixed installation CAT Ⅱ Local level mains appliances portable ...

Страница 29: ...3 23 EEC for Product Safety Compliance was demonstarted to the following specifications as listed in the official Journal of the European Communites En 55011 Class A Radiated and Conducted Emissions EN 50082 1 Immunity IEC 801 2 Electrostatic Discharge IEC 801 3 RF Radiated EN 61010 1 Safety requirements for electrical equipment for measurement control and laboratory use ...

Страница 30: ...or chemical cleaning agents BATTERY REPLACEMENT refer to Figure 4 1 Disconnect the test leads from any circuit under test and turn off meter 2 Remove the test leads from meter 3 Loosen the plastic screws on the battery cover 4 Remove battery cover from case bottom 5 Install a new battery after removing the original one 6 Assemble battery cover onto bottom case with screw driver FUSE REPLACEMENT re...

Страница 31: ...w Screw Battery Cover Battery Connector 9V Battery Figure 4 Screw Bottom Case Figure 5 Fuse1 1A 600v Battery Cover Screw Screw Fuse2 15A 600v Top Case Figure 4 Figure 5 Screw Screw Battery Cover Battery Connector 9V Battery Top Case Bottom Case Screw Screw Screw Battery Cover Fuse 1 1A 600V Fuse 2 15A 600V ...

Страница 32: ...N 109N Gift Box Meter Holster Tilt Battery 9V Alkaline Manual Test Leads Aligator Clip Temp Socket K Type Sensor 50BK RS 232 Cable with Adapter DB9M to DB25F Option Carrying Case Option Option 10 ACCESSORIES ...

Страница 33: ...APPA TECHNOLOGY CORP 9F 119 1 Pao Zong Rd Shin Tien Taipei 23115 Taiwan R O C P O Box 12 24 Shin Tien Taiwan Tel 886 2 29178820 Fax 886 2 29170848 E mail info appatech com Http www appatech com ...

Отзывы: