Solution
| Temporary solution- Disable
[Initialize Self-Test] (old UI version) or
[p-comb test] (new UI version) in the
Maintenance-Hardware Adjustment to
turn off the plastic comb detecting
function.
| Final solution- After disassembling
the machine, adjust the height of the
sensors of plastic comb hook, so that
it can detect the grabbing of plastic
comb normally, then fasten the
sensors.
TroubleShooting
Problem
The instrument cannot detect the
grabbing of plastic comb normally.
(Magnetic head will grab comb, but
show error message. Or will not grab
comb, but run normally.)
The plastic comb is not grasped firmly
and hits the shield plate, knocking off
the comb grabbing hook
Reagent spill in the machine should be cleaned up
immediately, otherwise the surface of the machine
will corrode. Clean the spill area with a dustless
cloth dampened in 80% ethanol.
Install the comb grabbing hook back again.
1. Power on the machine, press Maintenance, then
Replace Magnet Head.
2. The position of the magnet head will rise, and the
comb grabbing hook will open. Turn off the machine
at this time.
3. Unscrew the magnet head, and install the comb
grabbing hook and the corresponding screw back to
the hole.
4. For easier operation, to install the hook on the
left, may need to open the back cover of the
instrument.
5. After installing the hook, reinstall the magnet
head. Turn on the machine and check whether the
initialization self-tests are passed. Verify whether the
plastic comb grabbing is normal.
16