Plate

Position

Plate

Type

Plate

ID

Content 

Reagent volume for each

well

1

2

3

4

5

96 well

plate 

96 well

plate 

96 well

plate 

96 well

plate 

96 well

plate with

comb

Bind

EtOH

EtOH

Elution

Comb

-

Elution Solution

80% Ethanol

80% Ethanol

Sample

Lysis/Binding Solution

(EtOH added)**

-

85 

µ

L

500 

µ

L

500 

µ

L

200 

µ

L

520 

µ

L

*Check 96 well plate integrity before loading reagents, to avoid leakage of certain wells. 
**Always mix the bottle of Lysis/Binding Solution well right before pouring into reagent reservoir, which is
used to load plates, as the magnetic beads can precipitate and compromise results. Also vortex the
reagent reservoir for Lysis/Binding Solution before adding more Lysis/Binding Solution. 

C. Automated viral Total NA isolation on Apostle MagTouch 1000
Automation Platform 

1. Initialize the instrument, then press the [Quick Start] button. 

2. Select the built-in [viral RNA 2.0] program and place the plates into the Apostle
MagTouch 1000 as indicated on the instrument display. Make sure that the plate is
placed steadily in the correct orientation. 

Ensure that the A1 well is in the upper

right corner of the plate holder on turntable.

3. After placing all the plates, press the      icon on the interface to start the
program.

4. When the program is complete, collect the purified viral RNA from the elution
plate. 

5. For long term storage, store the samples at -80 C.

o

12

Содержание MagTouch 1000

Страница 1: ...Apostle MagTouch 1000 Nucleic Acids Extraction Automation System USER MANUAL Fast and efficient...

Страница 2: ...h a nucleic acids isolation automation system and Apostle MiniEnrich a high resolution DNA size enrichment technology using a magnetic nano platform Apostle responds to the COVID 19 pandemic to help o...

Страница 3: ...shipping box and place on the designated area 2 Remove and keep all the tapes and foam cushions used for transportation Figure 1 Do NOT connect power before removing the transportation fixing screw 5...

Страница 4: ...are stuck out If not wide open use pipette tip to open the wells Figure 4 refer to operation checkpoint 1 3 Press Initialization button on the top left corner to initialize the machine 5 Wait for the...

Страница 5: ...er 2 in the instrument through the plate loading window Ensure A1 well is in the upper right corner 14 Peel off the protection film on one Apostle EtOH Plate Figure 6 16 Press the Plate position 3 but...

Страница 6: ...d into the plate Figure 9 27 Place the comb and empty plate on the plate holder 5 in the instrument through the plate loading window Ensure A1 well is in the upper right corner 28 Close the plate load...

Страница 7: ...te position 4 button on the screen Figure 12 37 Open the plate loading window gently 38 Take out the Elution Plate from the plate loading window 39 Seal the Elution Plate with protection film Figure 1...

Страница 8: ...open as shown Figure 1 use pipette tip to make sure the wells are wide open 08 1 Figure 1 Figure 2 Checkpoint 2 After the start of the program check if the comb gripper successfully grips comb during...

Страница 9: ...t lifted manual movement of the plate will be limited and does not open up the wells Only adjust the plate when it is lifted up from the turntable Otherwise the turntable may be damaged Operation Chec...

Страница 10: ...er Take out the lysis binding plate and comb then move the plate forward backward to the left and right so that the comb can make the wells wide open Move the comb up and down several times and gently...

Страница 11: ...le MagTouch 1000 Apostle 96 well plates and 96 tip comb micropipettes 1 mL 200 L 20 L tips Adjustable single and multi channel Liquid reservoirs Vortex or shaker Ethanol 200 proof DNase RNase free wat...

Страница 12: ...sults Also vortex the reagent reservoir for Lysis Binding Solution before adding more Lysis Binding Solution C Automated viral Total NA isolation on Apostle MagTouch 1000 Automation Platform 1 Initial...

Страница 13: ...ide the machine immediately clean the area with a dustless cloth dampened into 80 ethanol Frequency when needed Fasten Screws Check whether the magnet head screw turntable screws and screws of the fix...

Страница 14: ...lization Turn off the machine Sterilization Frequency before every run Turn off the machine for at least 15 minutes between each shift Frequency between each shift Close the monitor window and the pla...

Страница 15: ...iation get directly on the skin The glass door of the instrument can block UV leakage Please keep the door closed when using the UV lamp If the door or glass of the door is damaged please contact the...

Страница 16: ...g off the comb grabbing hook Reagent spill in the machine should be cleaned up immediately otherwise the surface of the machine will corrode Clean the spill area with a dustless cloth dampened in 80 e...

Страница 17: ...position of the coupling 17 Temporary solution For the old software version this situation may occur randomly after the instrument runs for a long time Shut down the machine for a period of time and t...

Страница 18: ...e wells of the plate causing liquid leakage Temporary solution Users should pay attention before use Before adding liquid observe whether there are obvious holes at the bottom of the plate and observe...

Страница 19: ...ad will be blocked Straighten the screw rod and then turn on the machine to confirm whether the problem is solved There is abnormal noise when the instrument is turned on or initialized and there is n...

Страница 20: ...d the positions of the other plates are deviated Contact the supplier After the machine has been running for long periods of time the position of the parts is misaligned The technician hits or touches...