MAINTENANCE AND RECALIBRATION 

µCache Maintenance 

Make sure that the latest version of software is installed for the mobile app and the latest version of firmware is 
installed on the µCache. Use the app store for your operating system to confirm that you are using the latest 
version of Apogee Connect. The firmware version can be checked in the Settings page in the app while connected 
to the µCache.

 

The µCache unit should be kept clean and free of debris. 

If the housing is opened for any reason, extra care should be taken to ensure that the gasket and seating are clean 
and the interior stays free of moisture. The screws must be tightened until firm to create a weather-tight seal. 

Steps to Replace the µCache Battery 

1.

 

Use a Philips screwdriver to remove the screws from the battery cover. 

2.

 

Remove the battery cover. 

3.

 

Remove the used battery. 

4.

 

Place a fresh battery in its place aligning the positive terminal with the + label on the board. 

5.

 

Make sure the gasket and seating are clean. 

6.

 

Replace the battery cover. 

7.

 

Use a Philips screwdriver to replace the screws. 

Sensor Maintenance and Recalibration 

Moisture or debris on the diffuser is a common cause of low readings. The sensor has a domed diffuser and 
housing for improved self-cleaning from rainfall, but materials can accumulate on the diffuser (e.g., dust during 
periods of low rainfall, salt deposits from evaporation of sea spray or sprinkler irrigation water) and partially block 
the optical path. Dust or organic deposits are best removed using water or window cleaner and a soft cloth or 
cotton swab. Salt deposits should be dissolved with vinegar and removed with a soft cloth or cotton swab. Never 
use an abrasive material or cleaner on the diffuser. 

Although Apogee sensors are very stable, nominal accuracy drift is normal for all research-grade sensors. To 
ensure maximum accuracy, we generally recommend sensors are sent in for recalibration every two years, 
although you can often wait longer according to your particular tolerances. 

See individual sensor product manuals for more sensor-specific maintenance and recalibration information. 

 

 

Содержание mCACHE

Страница 1: ...APOGEE INSTRUMENTS INC 721 WEST 1800 NORTH LOGAN UTAH 84321 USA TEL 435 792 4700 FAX 435 787 8268 WEB APOGEEINSTRUMENTS COM Copyright 2020 Apogee Instruments Inc OWNER S MANUAL CACHE Rev 28 Oct 2020...

Страница 2: ...icate of Compliance 3 Introduction 4 Sensor Models 5 Specifications 5 Quick Start Guide 6 Cable Connectors 8 Deployment and Installation 9 Maintenance and Recalibration 11 Troubleshooting and Customer...

Страница 3: ...electrical and electronic products with respect to the restriction of hazardous substances Please be advised that based on the information available to us from our raw material suppliers the products...

Страница 4: ...also make live data measurements when connected to a mobile device The mobile app displays the data and allows the user to record samples in the app and download and export them Data logging is set up...

Страница 5: ...Battery Life 1 year w 10 second sampling interval and an average of 5 minutes daily connected time 2 years w 60 second sampling interval and an average of 5 minutes daily connected time Operating Env...

Страница 6: ...ect your sensor model and specify custom calibration factors if necessary You can also rename the Cache you want Hit ENTER 4 Your Cache is now shown on the app s main display with live readings Click...

Страница 7: ...3 seconds 2 Click on Collect Logs icon 3 Select Append to existing to add to an existing dataset already on your phone or Create New to start creating a new data set 4 Confirm Start and End date match...

Страница 8: ...we remove the unused pigtail lead colors at the datalogger end of the cable If a replacement cable is required please contact Apogee directly to ensure ordering the proper pigtail configuration Align...

Страница 9: ...or spot check measurements and through the built in logging feature To accurately measure incoming radiation the sensor must be level For this purpose each sensor model comes with a different option f...

Страница 10: ...Please Note Do not let the Cache dangle...

Страница 11: ...e the gasket and seating are clean 6 Replace the battery cover 7 Use a Philips screwdriver to replace the screws Sensor Maintenance and Recalibration Moisture or debris on the diffuser is a common cau...

Страница 12: ...p to 100 m of cable All Apogee sensors use shielded twisted pair cable to minimize electromagnetic interference For best measurements the shield wire must be connected to an earth ground This is parti...

Страница 13: ...factory The warranty does not cover equipment that has been damaged due to the following conditions 1 Improper installation or abuse 2 Operation of the instrument outside of its specified operating r...

Страница 14: ...nship Apogee Instruments will repair or replace the items free of charge If it is determined that your product is not covered under warranty you will be informed and given an estimated repair replacem...

Отзывы: