7. pH Measurement

7.1   How to Take pH measurements
     Short press 

 to turn on the tester. Rinse the probe in 

pure water, shake off excess water. Insert the probe in your 
sample solution, make a quick stir and hold still. Record the 
reading when it is stabilized

 

 comes up and stays on 

screen). Or if you turn on the Auto-Hold function (refer to 
Section 13.2 (2)), the reading will be automatically locked 
when it's stable. Short press 

 to cancel the Auto-Hold 

and keep measuring.

7.2  Pure Water pH Measurement
     When testing pH of pure water such as tap water, drinking water, RO water and distilled water, it 

will take longer for the readings to get fully stabilized (typically 1-5 minutes). Please be patient. 
If still not working, add Apera 3M KCL (AI1107) to your pure water at the ratio of 1:1000 (e.g. 1 ml 
KCL to 1000 ml water) to accelerate stabilization while minimizing pH change. If the accuracy is 
not meeting your requirement, please contact us at [email protected] to find the specialized 
meter designed for pure water pH test.

8. Conductivity Calibration

8.1  How to calibrate

8.1.1  Power on the tester. Press 

 to switch to conductivity (Cond) measurement mode. Rinse the 

probe in pure water and shake off excess water.

8.1.2  Pour conductivity calibration solutions into the corresponding calibration vials to about half 

volume.

8.1.3  Long press 

 to enter calibration mode (screen turns green).

8.1.4  Dip the probe into 1413 μS conductivity calibration solution, make a quick stir and hold still. 

When the reading is fully stabilized (

 stays on screen), short press 

 to finish 1st point 

calibration. Then the tester returns to measurement mode and   will appear at the bottom left, 
indicating a successful 1st point conductivity calibration.

8.1.5  If needed (your estimated sample conductivity level is greater than 2 mS or 2000 μS), insert 

the probe into the 12.88 mS conductivity calibration solution. Follow the steps in 8.1.3 to 8.1.4 to 
complete the 2nd point of calibration using the 12.88 mS standard solution. 

 will appear next 

to  , indicating a successful 2-point conductivity calibration.

Содержание ZenTest PC60-Z

Страница 1: ...1 PC60 Z Smart Water Tester pH Cond TDS Salinity Resistivity ORP Temp User Manual IP67 ISO 9001 2015 APERA INSTRUMENTS LLC aperainst com...

Страница 2: ...ure below Users can remove the shield when cleaning the sensor and put it back on after cleaning 1 Introduction 3 2 What s in the Kit 4 3 Keypad Functions 4 4 Preparation before Use 5 5 Battery Replac...

Страница 3: ...lable on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mo...

Страница 4: ...ong press to turn off the tester short press to turn on off backlight 1 In measurement mode short press to switch parameter pH Cond TDS Sal Res 2 In measurement mode long press to turn on off Bluetoot...

Страница 5: ...ionized water RO water or tap water is the alternative then shake off excess water 5 3 Perform calibration For pH calibration tutorial refer to Section 6 for conductivity calibration refer to Section...

Страница 6: ...tely after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH...

Страница 7: ...eck whether pH probe is damaged or broken If so please replace with a new one 2 T h e g l a s s b u l b o r j u n c t i o n i s s e ve re l y contaminated Please use a soft brush with soap water to cl...

Страница 8: ...at info aperainst com to find the specialized meter designed for pure water pH test 8 Conductivity Calibration 8 1 How to calibrate 8 1 1 Power on the tester Press to switch to conductivity Cond meas...

Страница 9: ...The meter cannot recognize the conductivity standard solutions 1 Make sure the probe is fully submerged in the solution 2 Check if the standard solution is expired or contaminated 3 Check if the condu...

Страница 10: ...1XXX ppm c The TDS and Salinity values are converted from the conductivity values via a certain conversion factor TDS and conductivity is linear related and its conversion factor is 0 40 to 1 00 Adju...

Страница 11: ...vity sensor 3 If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes over time by its nature Just rinse them off and keep...

Страница 12: ...the meter is turned off long press to enter parameter setting short press to switch P01 P02 P14 Short Press parameter flashes short press to adjust parameter short press to confirm Short press to exit...

Страница 13: ...or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to set the meter to its def...

Страница 14: ...the probe s position 5 Screw on the probe ring tightly Soak the probe in 3M KCL for 5 15 minutes Then perform calibration before testing The replacement probes that are compatible with PC60 Z PC60 DE...

Страница 15: ...o 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible damage of the probe and it s within th...

Страница 16: ...d void the user s authority to operate the equipment This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits a...

Отзывы: