6. pH Calibration

6.1  Short press 

 to turn on the meter; rinse the probe in pure water. Shake off excess water.

6.2 Pour buffer solutions into the corresponding calibration vials (to about half volume of the vials). 
6.3 Long press 

 to enter calibration mode (screen turns green); 

6.4 Insert the probe into pH 7.00 buffer solution, make a quick 

stir, and hold still. When the reading is stabilized (  stays on 
screen), short press 

 to finish 1st point calibration. Then the 

tester returns to measurement mode, icon 

 will appear at the 

bottom left, indicating a successful 1st point pH calibration.

6.5 To calibrate 2nd point, use 4.00 pH buffer and repeat Step 6.3 to 

6.4 (Do NOT turn off the tester after you finish pH 7 calibration). 

 will display next to 

, indicating a successful 2-point pH 

calibration (low and middle points). 

6.6  If necessary, calibrate 3rd point using 10.01 standard pH buffer and repeat Step 6.3 to 6.4, 

 

will show up next to 

 and 

, indicating a successful 3-point calibration (high, low, and middle 

points).

6.7  Notes about Calibration

1)  The 1st point calibration must be 7.00 pH. Perform the 2nd and 3rd point calibrations (4.00, 

10.01, 1.68, or 12.45) immediately after the 1st point calibration is finished. Do NOT turn off the 
meter before you calibrate 2nd or 3rd point. Otherwise, you will need to restart the calibration 
process with 7.00 pH first. 

2)  The pH 4.00 and 7.00 buffer solutions poured into the calibration vials can be used for up 

to 10 times as long as they are not contaminated and the bottles are capped when not in 
use. pH 10.01 can only be used for up to 5 times as it will lose its accuracy much faster. After 
that, replace the buffer solutions in the calibration vials with new ones to keep the accuracy. 
Keeping the freshness and cleanliness of calibration buffers is essential for accurate pH 
measurement.

3) The tester can perform 1 to 3 points of automatic calibration and can recognize 5 types of pH 

standard solutions. For details, please refer to the following table:

Calibration

USA Series

NIST Series

Indication 

icon

Recommended

1-point

7.00 pH

6.86 pH

Accuracy requirement 

≥ 0.1 pH

2-point

Option A

1st pt: 7.00 pH

2nd pt: 4.00 pH 

or 1.68 pH

Option A

1st pt: 6.86 pH

2nd pt: 4.01 pH or 

1.68 pH

 

 

Range < 7.00 pH

Option B

1st pt: 7.00 pH

2nd pt: 10.01 pH 

or 12.45 pH

Option B

1st pt: 6.86 pH

2nd pt: 9.18 pH or 

12.45 pH

 

 

Range >7.00 pH

3-point

1st pt: 7.00 pH

2nd pt: 4.00 or 1.68 pH

3rd pt: 10.01 or 12.45 pH

1st pt: 6.86 pH

2nd pt: 4.01 or 1.68 pH

3rd pt: 9.18 pH or 12.45 pH

 

 

Range: 0 to 14.00 pH

Содержание ZenTest PC60-Z

Страница 1: ...1 PC60 Z Smart Water Tester pH Cond TDS Salinity Resistivity ORP Temp User Manual IP67 ISO 9001 2015 APERA INSTRUMENTS LLC aperainst com...

Страница 2: ...ure below Users can remove the shield when cleaning the sensor and put it back on after cleaning 1 Introduction 3 2 What s in the Kit 4 3 Keypad Functions 4 4 Preparation before Use 5 5 Battery Replac...

Страница 3: ...lable on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mo...

Страница 4: ...ong press to turn off the tester short press to turn on off backlight 1 In measurement mode short press to switch parameter pH Cond TDS Sal Res 2 In measurement mode long press to turn on off Bluetoot...

Страница 5: ...ionized water RO water or tap water is the alternative then shake off excess water 5 3 Perform calibration For pH calibration tutorial refer to Section 6 for conductivity calibration refer to Section...

Страница 6: ...tely after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH...

Страница 7: ...eck whether pH probe is damaged or broken If so please replace with a new one 2 T h e g l a s s b u l b o r j u n c t i o n i s s e ve re l y contaminated Please use a soft brush with soap water to cl...

Страница 8: ...at info aperainst com to find the specialized meter designed for pure water pH test 8 Conductivity Calibration 8 1 How to calibrate 8 1 1 Power on the tester Press to switch to conductivity Cond meas...

Страница 9: ...The meter cannot recognize the conductivity standard solutions 1 Make sure the probe is fully submerged in the solution 2 Check if the standard solution is expired or contaminated 3 Check if the condu...

Страница 10: ...1XXX ppm c The TDS and Salinity values are converted from the conductivity values via a certain conversion factor TDS and conductivity is linear related and its conversion factor is 0 40 to 1 00 Adju...

Страница 11: ...vity sensor 3 If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes over time by its nature Just rinse them off and keep...

Страница 12: ...the meter is turned off long press to enter parameter setting short press to switch P01 P02 P14 Short Press parameter flashes short press to adjust parameter short press to confirm Short press to exit...

Страница 13: ...or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to set the meter to its def...

Страница 14: ...the probe s position 5 Screw on the probe ring tightly Soak the probe in 3M KCL for 5 15 minutes Then perform calibration before testing The replacement probes that are compatible with PC60 Z PC60 DE...

Страница 15: ...o 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible damage of the probe and it s within th...

Страница 16: ...d void the user s authority to operate the equipment This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits a...

Отзывы: