- 12 -

13. Parameter Settings

13.1  Table of Settings

Symbol

Parameter Setting Contents

Content

Factory 

Default

P1

Temperature Unit

˚C – ˚F

˚F

P2

Auto. HOLD

5-20 seconds – Off

Off

P3

Automatic Backlight Off

1-8 minutes – Off

1

P4

Automatic Power Off

10-20 minutes – Off

10

P5

pH Buffer Series Selection

USA – NIST

USA

P6

pH Resolution

0.1 – 0.01

0.01

P7

pH Calibration Reminder

H-hours D-Days (set up in ZenTest App)

/

P8

pH back to factory default

No – Yes

No

P9

Conductivity Reference 

Temperature

15˚C to 30˚C

25˚C

P10

Temp. Compensation Coefficient

0 to 9.99

2.00

P11

Conductivity Calibration Reminder

H-hours D-Days (set up in ZenTest App)

/

P12

Conductivity Back to Factory 

Default

No – Yes

No

P13

TDS Factor

0.40 to 1.00

0.71

P14

Salinity Unit

ppt – g/L

ppt

13.2 Parameter Settings Examples
1)  When the meter is turned off, long press 

 to enter parameter setting →short press 

 

to switch P01-P02… →P14. Short Press 

 , parameter flashes → short press 

 to adjust 

parameter → short press 

 to confirm →Short press 

 to exit parameter setting and go back 

to measurement mode.

2) 

Auto-Hold (P02)

 – You can set the auto-hold time from 5 to 20 seconds. For example, if 10 

seconds is set, when the reading is stable for more than 10 seconds, the reading will be 
automatically locked, and the HOLD icon will be displayed. Short press 

 again to cancel the 

lock. When the setting is "Off" , the auto-hold function is turned off, the reading can only be 
manually locked. Short press 

 to lock or unlock readings while measuring. 

3) 

Auto. Backlight (P03

) – You can set the automatic backlight time for 1 to 8 minutes. For example, 

if 3 minutes is set, the backlight will turn off automatically after 3 minutes; when the “Off” is set, 
the auto. backlight function will be turned off, and short press 

 to manually turn the backlight 

on or off.

Содержание ZenTest PC60-Z

Страница 1: ...1 PC60 Z Smart Water Tester pH Cond TDS Salinity Resistivity ORP Temp User Manual IP67 ISO 9001 2015 APERA INSTRUMENTS LLC aperainst com...

Страница 2: ...ure below Users can remove the shield when cleaning the sensor and put it back on after cleaning 1 Introduction 3 2 What s in the Kit 4 3 Keypad Functions 4 4 Preparation before Use 5 5 Battery Replac...

Страница 3: ...lable on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mo...

Страница 4: ...ong press to turn off the tester short press to turn on off backlight 1 In measurement mode short press to switch parameter pH Cond TDS Sal Res 2 In measurement mode long press to turn on off Bluetoot...

Страница 5: ...ionized water RO water or tap water is the alternative then shake off excess water 5 3 Perform calibration For pH calibration tutorial refer to Section 6 for conductivity calibration refer to Section...

Страница 6: ...tely after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH...

Страница 7: ...eck whether pH probe is damaged or broken If so please replace with a new one 2 T h e g l a s s b u l b o r j u n c t i o n i s s e ve re l y contaminated Please use a soft brush with soap water to cl...

Страница 8: ...at info aperainst com to find the specialized meter designed for pure water pH test 8 Conductivity Calibration 8 1 How to calibrate 8 1 1 Power on the tester Press to switch to conductivity Cond meas...

Страница 9: ...The meter cannot recognize the conductivity standard solutions 1 Make sure the probe is fully submerged in the solution 2 Check if the standard solution is expired or contaminated 3 Check if the condu...

Страница 10: ...1XXX ppm c The TDS and Salinity values are converted from the conductivity values via a certain conversion factor TDS and conductivity is linear related and its conversion factor is 0 40 to 1 00 Adju...

Страница 11: ...vity sensor 3 If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes over time by its nature Just rinse them off and keep...

Страница 12: ...the meter is turned off long press to enter parameter setting short press to switch P01 P02 P14 Short Press parameter flashes short press to adjust parameter short press to confirm Short press to exit...

Страница 13: ...or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to set the meter to its def...

Страница 14: ...the probe s position 5 Screw on the probe ring tightly Soak the probe in 3M KCL for 5 15 minutes Then perform calibration before testing The replacement probes that are compatible with PC60 Z PC60 DE...

Страница 15: ...o 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible damage of the probe and it s within th...

Страница 16: ...d void the user s authority to operate the equipment This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits a...

Отзывы: