Apera Instruments SX811-BS Скачать руководство пользователя страница 8

 

- 5 - 

Table – 1 Keypad operations and descriptions 

Keypad 

Operations 

Descriptions 

 

Short press 

l

 Press this key to turn on or turn off the meter. 

 

Short press 

Select measurement parameters:

 

 

l

 pH meter:           

            , 

Long press 

l

 In measurement mode, press this key to enter main menu. 

 

Short press 

l

 In measurement mode, press this key to enter in calibration       

mode, 

l

 In recall mode (RM), press this key to return to measurement 

mode, 

l

 Cancel any operation to return to measurement mode. 

 

Short press 

l

 In measurement mode, press this key to turn on or turn off 

backlight, 

l

 In calibration mode, press this key to conduct calibration, 

l

 In main menu, press this key to enter submenu, 

l

 In submenu, press this key to enter parameter set-up     

mode, 

l

 In parameter set-up mode, press this key to confirm parameters. 

Long press 

l

 In pH measurement mode, press and hold this key to change the       

resolution repeatedly:

 

0.01→ 0.1pH 

 

Short/ 

long press 

 

l

 In the mode of manual temperature compensation (MTC), when press     

and hold this key, the temperature value flashes, then press this key to 
change the temperature value, and press            to confirm, 

l

 In measurement mode, press          to store the measuring value,     

press          to recall the stored measuring value,   

l

 In recall mode (RM), short press this key to change the   

storage  numberings,  press  and  hold  this  key  to  change  the  number 
quickly, 

l

 In the main menu and submenu mode, press this key to change the   

numbering of the main menu and the submenu,   

l

 In the parameter set-up mode, press this key to select parameters.   

 
 

3.3  Display mode 

3.3.1  Reading stability display mode 

When the measuring value is stable, smiley icon          appears on LCD, see 

Diagram – 5. If          icon does not appear or flash, please do not get the 

reading value or make calibration until the measuring value is stable.       

 

Diagram - 3 

Содержание SX811-BS

Страница 1: ...SX811 BS Portable pH Meter for Meat Testing Instruction Manual APERA INSTRUMENTS LLC www aperainst com...

Страница 2: ...6 4 1 Information regarding pH Calibration 6 4 1 1 Standard Buffer Solution 6 4 1 2 3 Point Calibration 7 4 1 3 How often to calibrate 7 4 1 4 Calibration reminder 7 4 1 5 Check calibration time 8 4 1...

Страница 3: ...17 7 7 Data processing 17 8 What s in the box 17 9 Warranty 18 10 Appendix 18 10 1 Appendix I Parameter setup and factory default setup 18 10 2 Appendix II Abbreviation glossary 19 10 3 Appendix III S...

Страница 4: ...nced digital processing technology improvs meter responding time and accuracy There are reading stability display mode and automatic lock up display mode n IP57 waterproof operable under extreme condi...

Страница 5: ...do not scratch the stainless steel casing and blades of the electrode with sharp object or sand paper 2 Technical Specifications 2 1 Meter s Technical Specs Technical Specs pH Range 2 00 to 19 99 pH...

Страница 6: ...ng Life Reference System Electrolyte Polymer Shape of glass membrane Spear Range 0 to 14 pH Temperature Range 0 to 80 Connector BNC and RCA 3 Instrument Description 3 1 LCD Display 1 Measurement mode...

Страница 7: ...ture compensation 14 Calibration guide icon 15 Stability icon of readings 16 USB icon when this icon appears the meter connects the computer 17 Low battery icon when this icon appears please renew the...

Страница 8: ...ess lIn pH measurement mode press and hold this key to change the resolution repeatedly 0 01 0 1pH Short long press lIn the mode of manual temperature compensation MTC when press and hold this key the...

Страница 9: ...inutes the 2nd measuring value is stored See Diagram 8 the meter stores automatically eight measuring values When press key icon stops flashing and the meter stops automatic storage In automatic stora...

Страница 10: ...H 4 00 1 68 pH and 9 18 pH 7 00pH 4 00 1 68 pH and 10 01 pH 6 86pH 4 01 1 68pH 9 18 pH 0 to 14 00 pH 4 1 3 How often to calibrate The frequency that you need to calibrate your meter depends on the tes...

Страница 11: ...icon displays on LCD Press key to calibrate the meter End icon appears after calibration is done The 1st point calibration is finished In the meanwhile CAL2 will flash at the upper right corner and 4...

Страница 12: ...lue flashes Press key or key to adjust the measuring value to 1 60 then press key to calibrate the meter After calibration is done LCD at the top right shows blinking CAL2 icon indicating the meter en...

Страница 13: ...m parameter setup P1 6 please see Item 8 3 PU 2 icon displays at the right top of LCD please refer to Diagram 14 Note In parameter setup either PU 1 or PU 2 measurement mode can be selected but both c...

Страница 14: ...is to that of the calibration solution the more accurate the readings will be 6 Factory default setting For factory default setting please refer to parameter P1 6 Item 8 3 Per parameter P1 6 all calib...

Страница 15: ...g solution for 6 hours c For heavily polluted platinum surface on which there is oxidation film polish the platinum surface with toothpaste then wash it in pure water then submerge it in the soaking s...

Страница 16: ...flashes press to confirm 3 After confirm parameter press key to enter in P1 3 mode or press key to return to the measurement mode P1 3 Check the time and date of the last calibration 1 The time and d...

Страница 17: ...ult setting 2 When do not select Yes press key to return to the measurement mode 6 4 Submenu of basic parameter setup press key or key to switch P4 1 Adjust timing storage time 1 In mode P4 0 press ke...

Страница 18: ...2 After confirm the parameter press key to enter in P4 7 mode or press key to return to the measurement mode P4 7 Adjust date 1 Press key Month flashes then press key Date flashes then press key Year...

Страница 19: ...uter pH mV conductivity and dissolved oxygen are classified in the file Export press this key to export the stored value to Microsoft Excel file Exit press this key PC Link program exits from the comp...

Страница 20: ...downloaded to the computer pH and mV are sorted in the program 7 6 Storage during operation During operation press key to store or set timing storage The measuring information is downloaded to the com...

Страница 21: ...the products purchased by customers not the service life of the tester or probe 10 Appendix 10 1 Appendix I Parameter setup and factory default setup Modes Prompts Parameter setting items Abbreviation...

Страница 22: ...off Auto power off P4 5 Clear readings Clear readings P4 6 P4 7 P4 8 Others China China United States of America United States of America Nist Nist Off Off On On No No Yes Yes 10 3 Appendix III Self d...

Страница 23: ...20...

Страница 24: ...21 APERA INSTRUMENTS LLC Address 977 Old Henderson Rd Columbus Ohio 43220 Tel 1 614 285 3080 Email info aperainst com Website www aperainst com...

Отзывы: