- 10 - 

customer-defined calibration solution.   

 

4.4  Sample test 

1) Rinse pH electrode in distilled water, dry it, and dip it in sample solution. Stir the solution gently and let 

it stand still in the sample solution until the stable value and          icon appears on LCD, get the reading 

which  is  pH  value  of  sample  solution,  please  refer  to  Diagram-12  for  calibration  and  measurement 

process of pH meter. 

Diagram – 12 Calibration and measurement process of pH meter 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
2) pH measurement of pure water 
The meter is able to set up pH measurement mode of pure water with   
temperature compensation for pH value from parameter setup P1.5   
(please see Item 8.3). “

PU-1

” icon displays at the right top of LCD, please   

refer to Diagram – 13. 
 
3) pH measurement of pure water mixed with ammonia 
The meter is able to set up pH measurement mode of pure water mixed with 
ammonia with temperature compensation for pH value from parameter setup   
P1.6 (please see Item 8.3). “

PU-2

” icon displays at the right top of LCD, 

please refer to Diagram – 14. 

 

Note: In parameter setup, either “PU-1” or “PU-2” measurement mode 
can be selected, but both can not work at the same time. 

 

4) Self-diagnosis information 
During the process of calibration and measurement, the meter has self-diagnosis functions, indicating the 
relative information as below, please refer to chart – 6. 
 
 
 

Diagram - 13 

Diagram - 14 

Press          to turn 
    on the meter.     

Press          key to 

enter into the   

calibration mode. 

Submerge the 
electrode in pH7.00 
solution, press             
key when          icon     
appears. 

After the 1

st

 point 

calibration is done, 
press          key to 
enter into the 
measurement mode. 

 

 

 

Sample 

Test 

st

 point   

calibration 

n  d

  point 

calibration 

 

  

r   d

 point 

calibration 

 

After the 2

nd

 point 

calibration is done, 
press          key to 
enter into the 
measurement mode.

 

After the 3

rd

 point 

calibration is done, 
the meter enters into 
the measurement 
mode automatically.

 

Submerge the 
electrode in pH4.00 
solution, press               
key when          icon         
appears. 

Submerge the 
electrode in 
pH10.01 
solution,press         
key when          icon 
appears. 

 

Содержание SX811-BS

Страница 1: ...SX811 BS Portable pH Meter for Meat Testing Instruction Manual APERA INSTRUMENTS LLC www aperainst com...

Страница 2: ...6 4 1 Information regarding pH Calibration 6 4 1 1 Standard Buffer Solution 6 4 1 2 3 Point Calibration 7 4 1 3 How often to calibrate 7 4 1 4 Calibration reminder 7 4 1 5 Check calibration time 8 4 1...

Страница 3: ...17 7 7 Data processing 17 8 What s in the box 17 9 Warranty 18 10 Appendix 18 10 1 Appendix I Parameter setup and factory default setup 18 10 2 Appendix II Abbreviation glossary 19 10 3 Appendix III S...

Страница 4: ...nced digital processing technology improvs meter responding time and accuracy There are reading stability display mode and automatic lock up display mode n IP57 waterproof operable under extreme condi...

Страница 5: ...do not scratch the stainless steel casing and blades of the electrode with sharp object or sand paper 2 Technical Specifications 2 1 Meter s Technical Specs Technical Specs pH Range 2 00 to 19 99 pH...

Страница 6: ...ng Life Reference System Electrolyte Polymer Shape of glass membrane Spear Range 0 to 14 pH Temperature Range 0 to 80 Connector BNC and RCA 3 Instrument Description 3 1 LCD Display 1 Measurement mode...

Страница 7: ...ture compensation 14 Calibration guide icon 15 Stability icon of readings 16 USB icon when this icon appears the meter connects the computer 17 Low battery icon when this icon appears please renew the...

Страница 8: ...ess lIn pH measurement mode press and hold this key to change the resolution repeatedly 0 01 0 1pH Short long press lIn the mode of manual temperature compensation MTC when press and hold this key the...

Страница 9: ...inutes the 2nd measuring value is stored See Diagram 8 the meter stores automatically eight measuring values When press key icon stops flashing and the meter stops automatic storage In automatic stora...

Страница 10: ...H 4 00 1 68 pH and 9 18 pH 7 00pH 4 00 1 68 pH and 10 01 pH 6 86pH 4 01 1 68pH 9 18 pH 0 to 14 00 pH 4 1 3 How often to calibrate The frequency that you need to calibrate your meter depends on the tes...

Страница 11: ...icon displays on LCD Press key to calibrate the meter End icon appears after calibration is done The 1st point calibration is finished In the meanwhile CAL2 will flash at the upper right corner and 4...

Страница 12: ...lue flashes Press key or key to adjust the measuring value to 1 60 then press key to calibrate the meter After calibration is done LCD at the top right shows blinking CAL2 icon indicating the meter en...

Страница 13: ...m parameter setup P1 6 please see Item 8 3 PU 2 icon displays at the right top of LCD please refer to Diagram 14 Note In parameter setup either PU 1 or PU 2 measurement mode can be selected but both c...

Страница 14: ...is to that of the calibration solution the more accurate the readings will be 6 Factory default setting For factory default setting please refer to parameter P1 6 Item 8 3 Per parameter P1 6 all calib...

Страница 15: ...g solution for 6 hours c For heavily polluted platinum surface on which there is oxidation film polish the platinum surface with toothpaste then wash it in pure water then submerge it in the soaking s...

Страница 16: ...flashes press to confirm 3 After confirm parameter press key to enter in P1 3 mode or press key to return to the measurement mode P1 3 Check the time and date of the last calibration 1 The time and d...

Страница 17: ...ult setting 2 When do not select Yes press key to return to the measurement mode 6 4 Submenu of basic parameter setup press key or key to switch P4 1 Adjust timing storage time 1 In mode P4 0 press ke...

Страница 18: ...2 After confirm the parameter press key to enter in P4 7 mode or press key to return to the measurement mode P4 7 Adjust date 1 Press key Month flashes then press key Date flashes then press key Year...

Страница 19: ...uter pH mV conductivity and dissolved oxygen are classified in the file Export press this key to export the stored value to Microsoft Excel file Exit press this key PC Link program exits from the comp...

Страница 20: ...downloaded to the computer pH and mV are sorted in the program 7 6 Storage during operation During operation press key to store or set timing storage The measuring information is downloaded to the com...

Страница 21: ...the products purchased by customers not the service life of the tester or probe 10 Appendix 10 1 Appendix I Parameter setup and factory default setup Modes Prompts Parameter setting items Abbreviation...

Страница 22: ...off Auto power off P4 5 Clear readings Clear readings P4 6 P4 7 P4 8 Others China China United States of America United States of America Nist Nist Off Off On On No No Yes Yes 10 3 Appendix III Self d...

Страница 23: ...20...

Страница 24: ...21 APERA INSTRUMENTS LLC Address 977 Old Henderson Rd Columbus Ohio 43220 Tel 1 614 285 3080 Email info aperainst com Website www aperainst com...

Отзывы: