- 8 -
Paramet er Setting
Table of Settings
Parameter Setting
1)
When the meter is turned off, long press
to enter parameter setting →short press to switch
P01-P0
2… →P14. Short Press , parameter flashes → short press to adjust parameter →
short
press to confirm →Short press to exit parameter setting and go back to measurement
mode.
2)
Auto. Lock (P02)
– Users can set the auto lock time from 5 to 20 seconds. For example, if 10 seconds
is set, when the measured value is stable for more than 10 seconds, the measured value will be
automatically locked, and the HOLD icon will be displayed. Short press to release the lock. When
the setting is "Off" , the Auto. lock function is turned off, that is, the measured value can only be
manually locked. Short press to lock or unlock the measured value. The HOLD icon will be
displayed when reading is locked.
3)
Auto. Backlight (P03)
─
Users can set the automatic backlight time for 1 to 8 minutes. For example, if
3 minutes is set, the backlight will turn off automatically after 3 minutes; when the
“Off” is set, the auto.
backlight function will be turned off, and short press to manually turn the backlight on or off.
4)
Auto. Power off (P04)
─
The auto. power off time can be set to 10 to 20 minutes. For example, if 15
minutes is set, the meter will automatically shut down after 15 minutes if no operation; when
“Off” is set,
the auto. power off function will be turned off. Long press to manually shut down the meter.
Symbol
Parameter Setting Contents
Content
Factory
Default
P1
Temperature Unit
˚C – ˚F
˚F
P2
Select automatic lock
5-20 seconds
– Off
Off
P3
Automatic Backlight Off
1-8 minutes
– Off
1
P4
Automatic Power Off
10-20 minutes
– Off
10
P5
Conductivity Reference
Temperature
1
5 ˚C to 30 ˚C
2
5 ˚C
P6
Temp. Compensation Coefficient
0 to 9.99
2.00
P7
Conductivity Calibration Reminder
H-hours D-Days (set up in ZenTest App)
/
P8
Conductivity Back to Factory
Default
No
– Yes
No
P9
TDS Factor
0.40 to 1.00
0.71
P10
Salinity Unit
ppt
– g/L
ppt