background image

 

- 11 - 

 

Warrant y  

We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of 

charge, at option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of 

APERA INSTRUMENTS, LLC for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery. This limited 

warranty does NOT cover any damages due to: accidental damage, unauthorized repair, normal wear and tear, or external 

causes such as accidents, abuse, or other actions or events beyond our reasonable control. 

 

To claim for warranty or seek additional help, please go to 

support.aperainst.com

 and click 

New support ticket

Please provide the following information in the ticket: 

Your Name 

Phone Number 

Email 

Shipping Address 

Product Name 

Serial Number 

(on the back of the meter) 

Photo or screenshot of receipt 

Problems experienced 

 

After the ticket is submitted, one of our customer care specialists will be assisting you within 1 business day. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание AI3718

Страница 1: ... 1 SKU AI3718 EC60 Z Smart Multi Parameter Tester Conductivity TDS Salinity Resistivity Temp Instruction Manual APERA INSTRUMENTS LLC www aperainst com V1 03 ...

Страница 2: ... 10 Warranty 11 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the conductivity sensor before the product leaves factory It does NOT mean the product is used 2 The batteries are already preinstalled Just pull off the paper slip before using the tester When you replace the batteries make sure to follow the correct directi...

Страница 3: ...splay dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone following graphic guides Self Diagnosis Er1 Er6 icons Detailed problem analysis and solutions Parameter Setup Press buttons to set up except for P7 and P11 All parameters can be set up in Settings Alarm The s...

Страница 4: ...measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibration mode 2 In calibration mode short press to confirm calibration 3 In measurement mode when automatic lock is turned off short press to manually lock or unlock re...

Страница 5: ...re stained gently clean the electrode with a soft brush in warm water containing detergent or alcohol 3 Things needed in addition to what s in the box a Distilled or deionized water 8 16oz for rinsing the probe after each test b Tissue paper for drying the probe Conductivity Calibration How to calibrate 1 Press key to switch to conductivity measurement mode Cond Rinse the probe in distilled water ...

Страница 6: ...ffers we recommend that users replace new conductivity solutions after 5 to 10 times of use to keep the standard solution s accuracy Do NOT pour the used calibration solutions back into the solution bottles in case of contamination 5 Temperature compensation factor The default setting of the temperature compensation factor is 2 0 User can adjust the factor based on test solution and experimental d...

Страница 7: ...r bubbles on the surface of the black rods 2 Check if the conductivity electrode is contaminated If so please use a soft brush with warm water to clean up 3 Soak the probe in 12 88mS cm solution for 10 minutes then rinse with distilled water 4 If none of the above please contact us at info aperainst com or 1 614 285 3080 The calibration reminder is triggered It s time to perform a new conductivity...

Страница 8: ...automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 minutes when the Off is set the auto backlight function will be turned off and short press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut...

Страница 9: ...ing is cancelled in the ZenTest App the Er6 icon will disappear 6 Conductivity Back to Factory Default P8 Select Yes to recover instrument calibration to theoretical value This function can be used when instrument does not work well in calibration or measurement Calibrate and measure again after setting the instrument back to factory default Technical Specifications Conductivity Range 0 to 199 9 μ...

Страница 10: ...at is compatible with EC60 Z is EC60 DE Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6 Stable reading indicator Waterproof Rating IP67 floats on water Reading Lock HOLD Power DC3V AAA batteries 4 Bluetooth Signal Battery Life 200 Hours Low power reminder Backlight White Measurement Green Calibration Red Alarm Auto Power Off Automatically power off if no operation for 10 minutes Di...

Страница 11: ... to accidental damage unauthorized repair normal wear and tear or external causes such as accidents abuse or other actions or events beyond our reasonable control To claim for warranty or seek additional help please go to support aperainst com and click New support ticket Please provide the following information in the ticket Your Name Phone Number Email Shipping Address Product Name Serial Number...

Страница 12: ... 12 APERA INSTRUMENTS LLC Address 6656 Busch Blvd Columbus Ohio 43229 Tel 1 614 285 3080 Email info aperainst com Website www aperainst com ...

Отзывы: