  Complet e Kit  

 

 

 

 

 

 

 

 

 

Things to Know  Bef ore Use  

1) 

A few drops of water are added to the probe cap to keep the conductivity electrode in an activated state 
before  the  tester  leaves  factory.  Generally,  users  can  start  using  the  tester  directly.  For  conductivity 
electrode that hasn’t been used for a long time, users should soak the electrode in 12.88 mS calibration 
solution  for  5-10  minutes  or  in  tap  water  for  1  to  2  hours  before  use.  Rinse  the  electrode  in 
distlled/deionized water after each measurement. 

2) 

The  sensing  rod  of  Model  2301T-S  conductivity  electrode  is  coated  with  platinum  black  to  minimize 
electrode  polarization  and  expand  measuring  range.  The  platinum  black  coating  adopted  our  special 
processing technology, which improves the electrode performance and the firmness of the coating. If the 
black sensing rods are stained,  gently clean  the  electrode  with  a soft brush  in  warm water containing 
detergent or alcohol. 

3) 

Things needed in addition to what’s in the box: 

a) 

Distilled or deionized water (8-16oz) for rinsing the probe after each test 

b) 

Tissue paper for drying the probe 

 

Conduct ivit y Calibrat ion  

 

How to calibrate 

1) 

Press          key to switch to conductivity measurement mode (

Cond

). Rinse the probe in distilled water 

and dry it. 

2) 

Pour certain amount of 1413

μS/cm and 12.88mS/cm conductivity calibration solution into corresponding 

calibration bottles (to about half volume of the bottle). 

3) 

Long press          key to enter calibration mode, short press          to go back to measurement mode. 

4) 

Place the probe into 1413 

μS/cm conductivity calibration solution, shake it for a few seconds and allow it 

to stand still in the solution until a stable reading is reached. When            stays on the LCD screen, short 

press          key to complete 1

st

 calibration, the tester returns to measurement mode and indication icon           

            will appear at the bottom left of the LCD screen. 

Graph-2 

 

Содержание AI3718

Страница 1: ... 1 SKU AI3718 EC60 Z Smart Multi Parameter Tester Conductivity TDS Salinity Resistivity Temp Instruction Manual APERA INSTRUMENTS LLC www aperainst com V1 03 ...

Страница 2: ... 10 Warranty 11 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the conductivity sensor before the product leaves factory It does NOT mean the product is used 2 The batteries are already preinstalled Just pull off the paper slip before using the tester When you replace the batteries make sure to follow the correct directi...

Страница 3: ...splay dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone following graphic guides Self Diagnosis Er1 Er6 icons Detailed problem analysis and solutions Parameter Setup Press buttons to set up except for P7 and P11 All parameters can be set up in Settings Alarm The s...

Страница 4: ...measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibration mode 2 In calibration mode short press to confirm calibration 3 In measurement mode when automatic lock is turned off short press to manually lock or unlock re...

Страница 5: ...re stained gently clean the electrode with a soft brush in warm water containing detergent or alcohol 3 Things needed in addition to what s in the box a Distilled or deionized water 8 16oz for rinsing the probe after each test b Tissue paper for drying the probe Conductivity Calibration How to calibrate 1 Press key to switch to conductivity measurement mode Cond Rinse the probe in distilled water ...

Страница 6: ...ffers we recommend that users replace new conductivity solutions after 5 to 10 times of use to keep the standard solution s accuracy Do NOT pour the used calibration solutions back into the solution bottles in case of contamination 5 Temperature compensation factor The default setting of the temperature compensation factor is 2 0 User can adjust the factor based on test solution and experimental d...

Страница 7: ...r bubbles on the surface of the black rods 2 Check if the conductivity electrode is contaminated If so please use a soft brush with warm water to clean up 3 Soak the probe in 12 88mS cm solution for 10 minutes then rinse with distilled water 4 If none of the above please contact us at info aperainst com or 1 614 285 3080 The calibration reminder is triggered It s time to perform a new conductivity...

Страница 8: ...automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 minutes when the Off is set the auto backlight function will be turned off and short press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut...

Страница 9: ...ing is cancelled in the ZenTest App the Er6 icon will disappear 6 Conductivity Back to Factory Default P8 Select Yes to recover instrument calibration to theoretical value This function can be used when instrument does not work well in calibration or measurement Calibrate and measure again after setting the instrument back to factory default Technical Specifications Conductivity Range 0 to 199 9 μ...

Страница 10: ...at is compatible with EC60 Z is EC60 DE Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6 Stable reading indicator Waterproof Rating IP67 floats on water Reading Lock HOLD Power DC3V AAA batteries 4 Bluetooth Signal Battery Life 200 Hours Low power reminder Backlight White Measurement Green Calibration Red Alarm Auto Power Off Automatically power off if no operation for 10 minutes Di...

Страница 11: ... to accidental damage unauthorized repair normal wear and tear or external causes such as accidents abuse or other actions or events beyond our reasonable control To claim for warranty or seek additional help please go to support aperainst com and click New support ticket Please provide the following information in the ticket Your Name Phone Number Email Shipping Address Product Name Serial Number...

Страница 12: ... 12 APERA INSTRUMENTS LLC Address 6656 Busch Blvd Columbus Ohio 43229 Tel 1 614 285 3080 Email info aperainst com Website www aperainst com ...

Отзывы: