Reading is 

always 

slowly 

changing, 

won’t 

stabilize

.

 

Contaminated 

probe

 

Clean the probe with 

Apera’s cleaning solution or 

detergent water.

 

Clogged junction

 

Clean the probe with 

Apera’s cleaning solution, then soak 

it in 3M KCL soaking solution overnight.

 

Aged probe

 

Replace the probe.

 

Testing pH of low 

iconic strength 

solutions like tap 

water, drinking 

water, RO water

 

Be patient, wait for 2-5 minutes to reach a fully stabilized 

reading. If still not stabilizing, add 1ml of 3M KCL solution 

to 1000ml of test solution.

 

Display 

similar 

readings in 

any 

solutions or 

always 

display 7.0 

Broken probe

 

If you 

don’t find any visible damage of the probe and it’s 

within the 1-year probe warranty, contact your point of 

purchase for warranty fulfillment;

 

If there is visible damage 

or the probe is more than 1-year old, replace the probe.

 

Reading 

keeps 

jumping 

Probe is not fully 

submerged in the 

solution

 

Make sure the probe is fully immersed in the solution at 

least 1 inch deep.

 

Air bubbles 

around the sensor

 

Make a quick stir in the solution to remove air bubbles.

 

Probe is not 

properly 

connected, or the 

pin connector is 

broken.

 

Check the 

probe’s connector, make sure it’s not broken 

and is correctly connected. Align the probe and 

instrument correctly before plugging in.

 

Never force it.

 

Ensure that the probe connector is not exposed in the air 

too long.

 

Calibration 

is 

successful, 

but reading 

is not 

accurate 

Aged probe

 

Replace the probe.

 

Air bubbles 

around the sensor

 

Make a quick stir in the solution to remove air bubbles.

 

Clogged junction

 

Clean the probe with cleaning solution, then soak it in 3M 

KCL soaking solution overnight

 

Comparison with 

other testers, test 

strips, or drop 

tests

 

To compare with other testers, make sure to perform a 2-

point calibration for all testers in the same standards, then 

test a 3rd point. Whichever gives more accurate reading 

in the 3rd point standard is the most accurate one.

 

Test 

strips or drop 

tests’ accuracy is not comparable to pH 

meters’.

 

 

 

 

 

Out of range reminder for pH 

Out of range reminder for temperature 

 

Содержание 28APE190349/AI11010

Страница 1: ...GroStar GS1 Premium pH Pen Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...rket Since 1991 Apera instruments has been dedicated to providing hi tech accurate lab grade instruments and sensors GroStar s intelligent design reduces the guesswork so you can easily manage your cr...

Страница 3: ...s the level to which you should pour the soaking solution or sample solution For details of probe storage see Section 7 Premium pH Probe Built with low resistance lithium glass membrane with double ju...

Страница 4: ...e in 3M KCL soaking solution for 5 minutes before use 2 5 If the tester hasn t been used for a long time over 1 month please soak the probe in the 3M KCL soaking solution for at least 1 hour then cali...

Страница 5: ...up to the right of Calibrating pH 10 is usually not necessary unless your estimated target pH is greater than 8 0 pH 3 5 will disappear in 30 days after calibration reminding you to recalibrate the te...

Страница 6: ...1 If necessary you can manually hold lock the reading by short pressing Press it again to cancel the hold 5 2 Long press to switch temp unit between F and C 5 3 The tester will automatically power off...

Страница 7: ...store the probe in it Close the probe cap tightly with the O ring If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes...

Страница 8: ...as it will cause greater measurement error and damage to the probe Test your samples and perform calibration close to room temperature as much as possible 10 4 Never test oily liquids 10 5 Make sure...

Страница 9: ...ully immersed in the solution at least 1 inch deep Air bubbles around the sensor Make a quick stir in the solution to remove air bubbles Probe is not properly connected or the pin connector is broken...

Страница 10: ...alkaline batteries Backlight White measurement Green calibration Red errors Reading hold Manual Warranty Two years for the instrument one year for the probe pH probe Low resistance lithium glass membr...

Страница 11: ...proof of purchase as defined above and determines the Product is defective Apera may in its sole discretion either a repair the Product with new or refurbished parts or b replace the Product with a n...

Страница 12: ...D HAS NOT MADE ANY EXPRESS OR IMPLIED WARRANTY OR OTHER REPRESENTATION REGARDING THE PRODUCT AND DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TO THE EXTENT PERMITTED B...

Страница 13: ...12 APERA INSTRUMENTS Europe GmbH Wilhelm Muthmann Stra e 18 42329 Wuppertal Germany Contact info aperainst de Website www aperainst de Tel 49 202 51988998...

Отзывы: