3.3 

Hold 

 until screen turns green. The pen starts automatic 

calibration. Wait for "

Good

" to show up (in 10-15 seconds), 

indicating the calibration is completed, then the pen returns to 

measurement mode. 

 

Short press any key while calibrating (in green screen) to cancel 

calibration and return to measurement. 

 

3.4 

  icon will show up on the lower left corner of the screen 

indicating the pen is successfully calibrated. Repeat Step 3.2 to 3.3 

to calibrate pH 4 using 4.00 pH buffer, then 

 will show up next 

to 

You can continue to calibrate pH 10 by repeating Step 3.2 to 3.3 

using 10.01 pH buffer (sold separately), then 

 will show up to the right of 

 

Calibrating pH 10 is usually not necessary unless your estimated target pH is greater than 8.0 

pH. 

 

3.5 

 

 will disappear in 30 days after calibration, 

reminding you to recalibrate the tester. We recommend calibrating 

pH at least once a month to ensure accuracy. If you feel like the 

accuracy might be off, simply test the standard buffers (make sure 

the buffers are fresh and clean). If a discrepancy is fou

nd, then it’s 

time to calibrate again. 

3.6 

If the calibration fails, the screen will turn red. For details, see 

Section 11 Troubleshooting Guide. 

 

Always perform at least a 2-point pH calibration to ensure accuracy. Start 

with 7.00 pH, followed by 4.00 pH immediately. If you happened to turn off 

the pen before calibrating 4.00 pH, you need to start with 7.00 pH again 

after rebooting the pen, then 4.00 pH. 

 

Содержание 28APE190349/AI11010

Страница 1: ...GroStar GS1 Premium pH Pen Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...rket Since 1991 Apera instruments has been dedicated to providing hi tech accurate lab grade instruments and sensors GroStar s intelligent design reduces the guesswork so you can easily manage your cr...

Страница 3: ...s the level to which you should pour the soaking solution or sample solution For details of probe storage see Section 7 Premium pH Probe Built with low resistance lithium glass membrane with double ju...

Страница 4: ...e in 3M KCL soaking solution for 5 minutes before use 2 5 If the tester hasn t been used for a long time over 1 month please soak the probe in the 3M KCL soaking solution for at least 1 hour then cali...

Страница 5: ...up to the right of Calibrating pH 10 is usually not necessary unless your estimated target pH is greater than 8 0 pH 3 5 will disappear in 30 days after calibration reminding you to recalibrate the te...

Страница 6: ...1 If necessary you can manually hold lock the reading by short pressing Press it again to cancel the hold 5 2 Long press to switch temp unit between F and C 5 3 The tester will automatically power off...

Страница 7: ...store the probe in it Close the probe cap tightly with the O ring If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes...

Страница 8: ...as it will cause greater measurement error and damage to the probe Test your samples and perform calibration close to room temperature as much as possible 10 4 Never test oily liquids 10 5 Make sure...

Страница 9: ...ully immersed in the solution at least 1 inch deep Air bubbles around the sensor Make a quick stir in the solution to remove air bubbles Probe is not properly connected or the pin connector is broken...

Страница 10: ...alkaline batteries Backlight White measurement Green calibration Red errors Reading hold Manual Warranty Two years for the instrument one year for the probe pH probe Low resistance lithium glass membr...

Страница 11: ...proof of purchase as defined above and determines the Product is defective Apera may in its sole discretion either a repair the Product with new or refurbished parts or b replace the Product with a n...

Страница 12: ...D HAS NOT MADE ANY EXPRESS OR IMPLIED WARRANTY OR OTHER REPRESENTATION REGARDING THE PRODUCT AND DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TO THE EXTENT PERMITTED B...

Страница 13: ...12 APERA INSTRUMENTS Europe GmbH Wilhelm Muthmann Stra e 18 42329 Wuppertal Germany Contact info aperainst de Website www aperainst de Tel 49 202 51988998...

Отзывы: