3-Color LCD Screen 

 

White for measurement mode 

 

Green for calibration mode 

 

Red for calibration error 

 stands for stabilized 

reading 

L/M/H 

stands  for  successful 

calibration: 

is pH 4, 

is pH 7 

or 2.77 EC, 

is pH 10. The icon(s) 

will  disappear  after  30  days  to 
remind you for re-calibration. 

Probe Cap 

 

Water  droplets  are  added 
during  production  to  maintain 
the moisture of the probe. This 
is  normal  practice  and  should 
not  be  attributed  to  used 
product.

 

 

The  fill  line  shows  the  level  to 
which  you  should  pour  the 
soaking  solution  or  sample 
solution. 

 

For details of probe storage, see 
Section 7. 

Premium pH Probe 

 

Built  with  low-resistance  lithium  glass 
membrane with double-junction and blue 
gel  electrolyte,  ensuring  accuracy  and 
durability. 

 

The probe is replaceable, saving money in 
the long run. 

 

Short Press 

Long Press 

 

1. Power on 

2. Manual hold 

Press again to 

cancel hold.

 

Power off

 

 

In 

measurement 

mode, press to 

turn on/off the 

backlight. 

Start calibration 

 

In calibration 

mode, press to 

cancel 

calibration. 

Switch between 

°F and °C 

Sensor Shield 

Protects 

the 

sensitive 

pH 

glass 

membrane  from  accidental  damage. 
Remove it when cleaning the pH sensor. 

Содержание 28APE190349/AI11010

Страница 1: ...GroStar GS1 Premium pH Pen Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...rket Since 1991 Apera instruments has been dedicated to providing hi tech accurate lab grade instruments and sensors GroStar s intelligent design reduces the guesswork so you can easily manage your cr...

Страница 3: ...s the level to which you should pour the soaking solution or sample solution For details of probe storage see Section 7 Premium pH Probe Built with low resistance lithium glass membrane with double ju...

Страница 4: ...e in 3M KCL soaking solution for 5 minutes before use 2 5 If the tester hasn t been used for a long time over 1 month please soak the probe in the 3M KCL soaking solution for at least 1 hour then cali...

Страница 5: ...up to the right of Calibrating pH 10 is usually not necessary unless your estimated target pH is greater than 8 0 pH 3 5 will disappear in 30 days after calibration reminding you to recalibrate the te...

Страница 6: ...1 If necessary you can manually hold lock the reading by short pressing Press it again to cancel the hold 5 2 Long press to switch temp unit between F and C 5 3 The tester will automatically power off...

Страница 7: ...store the probe in it Close the probe cap tightly with the O ring If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes...

Страница 8: ...as it will cause greater measurement error and damage to the probe Test your samples and perform calibration close to room temperature as much as possible 10 4 Never test oily liquids 10 5 Make sure...

Страница 9: ...ully immersed in the solution at least 1 inch deep Air bubbles around the sensor Make a quick stir in the solution to remove air bubbles Probe is not properly connected or the pin connector is broken...

Страница 10: ...alkaline batteries Backlight White measurement Green calibration Red errors Reading hold Manual Warranty Two years for the instrument one year for the probe pH probe Low resistance lithium glass membr...

Страница 11: ...proof of purchase as defined above and determines the Product is defective Apera may in its sole discretion either a repair the Product with new or refurbished parts or b replace the Product with a n...

Страница 12: ...D HAS NOT MADE ANY EXPRESS OR IMPLIED WARRANTY OR OTHER REPRESENTATION REGARDING THE PRODUCT AND DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TO THE EXTENT PERMITTED B...

Страница 13: ...12 APERA INSTRUMENTS Europe GmbH Wilhelm Muthmann Stra e 18 42329 Wuppertal Germany Contact info aperainst de Website www aperainst de Tel 49 202 51988998...

Отзывы: