7. SETTING OUTPUT SIGNAL VALUES
By pushing
or
in the simulation mode (OUT), you can change the set value by the preset step
(parameter 9:
StEP
). By pressing SET and
at the same time, you set the output to the lower of the following
values: the upper limit of the simulated sensor or the value of parameter 8:
LiHi
; by pressing SET +
at the
same time, you set the output to the higher of the following values: the lower limit of the simulated sensor or
the value of parameter 7:
LiLo
. Also, the output signal can be set in the parameter programming mode
(parameter 10:
SEt
).
8. LIST OF MESSAGES AND ERRORS
^^^^
…..
upper segments of the display - the upper limit of the sensor range is exceeded or the sensor is
defective,
____
…..
lower segments of the display - the lower limit of the sensor range is exceeded or the sensor is
defective,
ConF
... the parameter configuration mode is entered,
CodE
... the mode for entering the password for access to the configuration parameters is entered,
Err
....... an incorrect password for access to the configuration parameters was entered,
----
....... initiation of the operating mode (input/output),
E-iH
..... the value of the current polarizing the output of the (thermo)resistance sensor is exceeded
(Ip > 1.5 mA)
E-iP
…..reverse polarization of the output of the (thermo)resistance sector (wrong connection to the
terminals),
200.0
. .. blinking set value - no current polarizing the output of the (thermo)resistance sensor,
(Ip < 25
µ
A)
E-rL
..... a short circuit in the output of the thermocouple sensor or the voltage sensor,
PoFF
... the calibrator was switched off (manually or automatically due to low battery voltage),
LbAt
…...battery voltage too low (replace batteries with new ones).