USER INSTRUCTION

 

 

 

 

 

TEMPERATURE CALIBRATOR 

AR915

 

 
 
 

 

Version 1.1.0

 

2015-06-12

 

Содержание AR915

Страница 1: ...S OFFICE 05 090 Raszyn ul Gałczyńskiego 6 Tel 48 22 853 48 56 48 22 853 49 30 48 22 101 27 31 E mail automatyka apar pl Website www apar pl USER INSTRUCTION TEMPERATURE CALIBRATOR AR915 Version 1 1 0 2015 06 12 ...

Страница 2: ...NCIPLES OF SAFE USE 2 2 GENERAL CHARACTERISTICS OF THE CALIBRATOR 2 3 TECHNICAL DATA 3 4 DIMENSIONS OF THE HOUSING AND DESCRIPTION OF CONNECTIONS AND EXTERNAL ELEMENTS 4 5 BUTTON FUNCTIONS 5 6 PROGRAMMING OF CONFIGURATION PARAMETERS 5 7 SETTING OUTPUT SIGNAL VALUES 7 8 LIST OF MESSAGES AND ERRORS 7 9 USER S NOTES 7 Please pay particular attention to the text marked with this sign The manufacturer ...

Страница 3: ...d a functional keyboard battery 2 x 1 5 V or rechargeable battery 2 x 1 2 V NiMH power supply AA type long operation on a fully charged set of batteries automatic power supply cutoff at low battery charging level automatic shutdown after a preset idle time possibility to test measurement devices and temperature sensors diagnostic functions facilitating detections such defects in tested circuits as...

Страница 4: ...nsor range 1 digit setting simulation Pt100 Ni100 2 C for Ip 200 µA and 3 C for Ip 200 µA J K E N 55 mV 0 3 2 kΩ 0 2 for sensor range and 1 5 Ωfor 0 3 2 kΩ S B R T 0 3 of sensor range Additional error for setting Pt100 Ni100 0 3 2kΩ 2 5 C or 1Ω nonlinearity Additional error for thermocouple inputs 2 C present only in automatic compensation of cold tip temperature Additional error from temperature ...

Страница 5: ...100 Ni100 termin als description 1 2 3 input Pt100 Ni100 Pt500 Pt1000 resistance 3 wire 1 3 input Pt100 Ni100 Pt500 Pt1000 resistance 2 wire 2 3 thermocouple input J K S B R T E N and 0 55 mV OUTPUT OUT Ip Pt100 Ni100 Ω TC mV termin als description 1 3 output Pt100 Ni100 resistance 1Ω 3 thermocouple output J K S B R T E N and 0 55 mV Ip resistance output polarizing current in simulations BATTERY I...

Страница 6: ...of detected connection for a resistance sensor 2 and 3 wire in measurements reference joint temperature for a thermocouple sensor in simulations and in measurements shows the calibrator software version 6 PROGRAMMING OF CONFIGURATION PARAMETERS press the CONF button until approx 2 s the display shows the following temporary message ConF or CodE if password protection is set i e parameter 15 Prot o...

Страница 7: ... 1 C 1 0 1 C 7 LiLo lower setting limit for SEt 00 0 3200 00 0 C 8 LiHi upper setting limit for SEt 00 0 3200 3200 9 StEP value change step SEt 0 5 200 0 1 0 C 10 SEt set value in the range of the sensor or LiLo LiHi 100 0 C 11 cALo zero point offset 50 0 50 0 C 0 0 C 12 cALG gain sensitivity 85 0 115 0 100 0 13 bLoc keyboard keys lock IN OUT and 10 SET oFF no lock inou lock IN and OUT SEt lock SE...

Страница 8: ...e display the lower limit of the sensor range is exceeded or the sensor is defective ConF the parameter configuration mode is entered CodE the mode for entering the password for access to the configuration parameters is entered Err an incorrect password for access to the configuration parameters was entered initiation of the operating mode input output E iH the value of the current polarizing the ...