![APAR AR234 Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/apar/ar234/ar234_user-instruction_2951921005.webp)
5
- additional for thermocouples
<2 °C (temperature of cold tips)
- additional from ambient temperature changes
< 0.005% of the input range /°C
An integrated temperature sensor
(measurement range:
-20 ÷ 70 °C)
Accuracy: ±0.5°C (in the range of -10 ÷ 70 °C)
±0.5 ÷ 1.5 °C (in the remaining range)
Resolution of measured temperature
0.1 °C
Communication interface for work with a computer
USB, drivers compatible with Windows XP/Vista/7/8/10
Measurement and data recording interval
programmable 10 s to 24 hours
Data storage memory
(non-volatile):
- internal
4MB, FLASH, file system FAT12, recording of up to 80,000
measurements
- external SD/MMC card
(connection with an ejector)
FAT16, FAT32, maximum size 2 GB, recommended size
≤ 1GB and FAT16
(recording up to 19 million
measurements)
Real time clock (RTC)
quartz, date, time, takes leap years into account)
Optical signaling
an LCD display, 3 LED’s: “READ/WRITE”, “Status”, 1 alarm
An LCD display
7-segment, 4 digits, 10 mm high
Power supply
(lithium battery)
3.6 V type AA (R6), 2,450 mAh, (e.g. SAFT LS14500)
Time of operation with a new battery (1)
up to 5 years (in temperatures 20 ÷ 30 °C)
Rated operating conditions
-20 to 70 °C, <100% RH (no condensation)
Operating environment
air and neutral gases, no dust
Housing
(wall-mounting)
material ABS UL94-V0, white
Protection rating
IP20 (no protection against ingress of water and
condensation)
Enclosure dimensions
80 x 80 x 25 mm
Operating position
any
Weight
approx. 100 g (with a battery)
Electromagnetic compatibility (EMC)
immunity: according to the PN-EN 61000-6-2:2002(U)
standard
emissivity: according to the PN-EN 61000-6-3:2002(U)
standard
(1)
−
operation time depends on the measurement period, the presence of an SD/MMC card, the operation mode of the LCD display, and the ambient
temperature:
−
5 years (recording period > 10 min., recording in internal memory, data copied only through the USB, LCD in the economy mode, 20-30 °C)
−
20 months (measurement period > 10 min., recording in internal memory, data copied only through the USB, LCD in the continuous mode,
20-30 °C)
−
7 months (measurement period 10 s, internal memory, data copied only through the USB, 20-30 °C)
−
1.5 year (measurement period > 10 min., recording on an SD/MMC card, LCD in the economy mode, 20-30 °C)
−
4 months (measurement period 10 s, recording on an SD/MMC card, 20-30 °C)
−
an unused SD/MMC card installed in the socket also reduces the operating time of a battery
−
it takes about 2 min. to transfer the entire content of the internal memory (4 MB) onto an SD/MMC card and it consumes about 1-2 mAh of the
battery’s capacity (tested on SanDisk and Kingston memory cards)
−
by using an accessory power supply device, the operating time of a new battery can be extended to about 8 years