Button function instructions
Function
Symbol
Operation instructions
Unlock
Unlock before adjusting operating parameters.
Run / Pause
Device starts running when power is on. Press "STOP" to standby and no high voltage
output. Press "RUN" again and the device will start working.
Adjust
cleaning cycle
CT
TH
→ +/-
Press CTTH first, then +/- to increase or decrease the cleaning cycle. The default cleaning
period is 15 days, and the adjustment range is 1-60 days. The cleaning function will alarm when
the countdown is 0 days. Power off and restart to restore the set time. The cleaning function
becomes invalid if the value is set to 0 days.
Adjust
frequency
FREQ → +/-
Press FREQ first
,
then press +/- to increase or decrease the frequency of positive and negative
ions output .
Adjust
ion balance voltage
IB
N
→ +/-
Press IB
N
first, then press + when the positive voltage on flat panel detector or discharge object
surface is high, press IB
N
first, then press - when the negative voltage on flat panel detector or
discharge object surface is high until the ion balance reaches the ideal status.
Confirm
CAL → OK
Note: Press CAL first, then press OK to confirm after adjusting the output parameter; otherwise,
false alarms are likely to occur.
Intelligent Control
Ion balance/
ion output frequency adjustable