"Transmission / Simulation" Function
D
NTA47271-300A2
179/184
D.3 "Transmission / Simulation" Function
Maximum rated voltage in common mode: 60 VDC or VAC.
D.3.1
DC Voltage
Range
Resolution
Measuring range
Accuracy / 1 year
Notes
+100 mV
1 µV
- 5 mV to + 100 mV
0.005% R + 2 µV
Output load min = 1 KOhm
+1 V
10 µV
- 5 mV to + 1 V
0.005% R + 8 µV
Output load min = 2 KOhm
+10 V
100 µV
- 100 mV to + 10 V
0.007% R + 80 µV
Output load min = 4 KOhm
+50 V
1 mV
- 100 mV to + 50 V
0.007% R + 0.5 mV
Output load min = 4 KOhm
Temperature coefficient < 5 ppm/°C from 0°C to 18°C and from 28°C to 50°C.
D.3.2
DC Current
Range
Resolution
Accuracy / 1 year
Notes
24 mA
1 µA
0.007% R + 0.8 µA
4-20 mA
1 µA
0.007% R + 0.8 µA
0-20 mA
1 µA
0.007% R + 0.8 µA
Temperature coefficient < 7 ppm/°C from 0°C to 18°C and from 28°C to 50°C.
The above specifications are given for the following configurations:
- CALYS 1500 active mode (+24 V ON)
Passive mode measurer (+24 V OFF).
- CALYS 1500 passive mode (+24 V OFF)
Active mode measurer (+24 V ON).
D.3.3
Resistance
Range
Resolution
Measuring range
Accuracy / 1 year
Notes
400
Ω
- 1 mA
(direct current)
10 m
Ω
1
Ω
to 400
Ω
0.006% R + 20 m
Ω
Iext of 0.1 mA / 1 mA
400
Ω
- 1 mA
(Pulsed current)
10 m
Ω
1
Ω
to 400
Ω
0.006% R + 30 m
Ω
Iext of 0.1 mA / 1 mA
400
Ω
- 4 mA
(direct current)
10 m
Ω
1
Ω
to 400
Ω
0.006% R + 20 m
Ω
Iext of 1 mA / 4 mA
400
Ω
- 4 mA
(Pulsed current)
10 m
Ω
1
Ω
to 400
Ω
0.006% R + 30 m
Ω
Iext of 1 mA / 4 mA
3600
Ω
(direct current)
100 m
Ω
10
Ω
to 3600
Ω
0.006% R + 100 m
Ω
Iext of 0.1 mA / 1 mA
3600
Ω
(Pulsed current)
100 m
Ω
10
Ω
to 3600
Ω
0.006% R + 200 m
Ω
Iext of 0.1 mA / 1 mA
Accuracy applies to a 4-wires connection.
Temperature coefficient < 5 ppm/°C from 0°C to 18°C and from 28°C to 50°C.
Содержание CALYS 1500
Страница 2: ...2 184 NTA47271 300A2 ...
Страница 3: ...NTA47271 300A2 3 184 CALYS 1500 Calibrateur de Process Multifonction Mode d emploi ...
Страница 15: ...Logiciel A NTA47271 300A2 15 184 ...
Страница 58: ...GENERALITES 58 184 NTA47271 300A2 CALYS 1500 Consigne entrée manuellement ...
Страница 95: ...NTA47271 300A2 95 184 CALYS 1500 Multifunction Process Calibrator Instruction manual ...