B.2.10 Pressure measurement
Open the MEASUREMENT MENU
Select the Pressure measurement function.
Select the unit (BAR, PSI, Pa,
Press ENTER.
Connect the pressure sensor on the right
connected or is defective, an error message will be displayed.
B.2.11 Modules With HART® Protocol
The HART® protocol (Highway-Addressable Remote Transducer) is a communications protocol over a direct current loop
(0-20 or 4-20mA). This protocol is used to without interfering with current measurement to both read measurements
carried out by the HART® transmitter and transmitter
The CALYS 1500 with related HART® module (ACL500) is used to carry out all of these operations.
To use the CALYS 1500 in HART mode, the unit must be connected to the current loop (across 24
channel V1) and be provided with the specific ACL500 modem (connected to the unit's side).
The calibrator provides the loop power supply via an internal 250
When an external loop power supply is used, a resistor from 230
loop power supply and transmitter.
Probe
connection
4-20 mA link with loop power
supply and HART® resistor
ON.
Measurement on channel V1
NTA47
MEASUREMENT MENU dialog box,
measurement function.
elect the unit (BAR, PSI, Pa, Atm, Kgcm2, cmHg, mmHg, inHgftH2O, inH2O).
Connect the pressure sensor on the right-hand side of the unit (see chapter, side connectors). If this is not
, an error message will be displayed.
Protocol
Addressable Remote Transducer) is a communications protocol over a direct current loop
This protocol is used to without interfering with current measurement to both read measurements
e HART® transmitter and transmitter-related information (brand, reference, etc..), and also to adjust it.
with related HART® module (ACL500) is used to carry out all of these operations.
in HART mode, the unit must be connected to the current loop (across 24
channel V1) and be provided with the specific ACL500 modem (connected to the unit's side).
es the loop power supply via an internal 250
Ω
series resistor.
When an external loop power supply is used, a resistor from 230
Ω
to 270
Ω
must be series connected with the external
HART® Transmitter
Measurement on channel V1
B
NTA47271-300A2
123/184
hand side of the unit (see chapter, side connectors). If this is not
Addressable Remote Transducer) is a communications protocol over a direct current loop
This protocol is used to without interfering with current measurement to both read measurements
related information (brand, reference, etc..), and also to adjust it.
with related HART® module (ACL500) is used to carry out all of these operations.
in HART mode, the unit must be connected to the current loop (across 24 V Out terminals from
channel V1) and be provided with the specific ACL500 modem (connected to the unit's side).
must be series connected with the external
Modem
HART®
AOIP ACL500
HART® Transmitter
Connexion modem
HART®
Содержание CALYS 1500
Страница 2: ...2 184 NTA47271 300A2 ...
Страница 3: ...NTA47271 300A2 3 184 CALYS 1500 Calibrateur de Process Multifonction Mode d emploi ...
Страница 15: ...Logiciel A NTA47271 300A2 15 184 ...
Страница 58: ...GENERALITES 58 184 NTA47271 300A2 CALYS 1500 Consigne entrée manuellement ...
Страница 95: ...NTA47271 300A2 95 184 CALYS 1500 Multifunction Process Calibrator Instruction manual ...