
ADVANCED OPERATION
140 / 164 NTA47265-300A2
CALYS 1000 / 1200
To enter calibration points in the table, use the
button.
Enter the values and press ENTER.
Use the following buttons to continue configuring a sensor.
To edit an existing calibration point,
To add a calibration point,
To delete a calibration point.
Between 1 and 4 calibration points can be entered per sensor.
These calibration points are used to calculate a polynomial c(T) of degree 0 to 3, giving the sensor's voltage (or
resistance) correction at temperature T.
In the specific case where a single calibration point is given, the behavior will differ according to whether the sensor is a
thermocouple or a thermometric resistance:
In the case of a thermocouple, the correction is a fixed voltage deviation.
In the case of a resistive probe, the correction made is an R0 correction.