GETTING STARTED
108 / 164 NTA47265-300A2
CALYS 1000 / 1200
Measuring DC voltage
B.2.1
Display the MEASUREMENT CONFIGURATION dialogue box:
Select the
Vdc
measurement function then the correct measurement range using the function and navigation
keys.
Confirm with ENTER.
The following ranges are available:
Range
100 mV
1 V
10 V
50 V
Auto
Resolution
1 µV
10 µV
100 µV
1 mV
---
Input impedance
> 10 k
Ω
> 10 k
Ω
1 M
Ω
1 M
Ω
---
The DC voltage to be measured is connected between the V and COM terminals.