GETTING STARTED
114 / 164 NTA47265-300A2
CALYS 1000 / 1200
Pulse counting
B.2.7
Display the MEASUREMENT CONFIGURATION dialogue box:
Select the
Counting
measurement
function, then the “type of input”
Signal or Hard Contact
using the function
and navigation keys.
Enter the counting time using the alphanumeric keys.
Confirm with ENTER.
Connection is made between the Hz and COM terminals.
The input signals are converted into a logic signal as in the case of signal frequency or hard contact measurement.
If you enter a zero counting time, the counting lasts indefinitely.
To initiate counting, confirm with the
Start
function key (“F3”).
To stop counting without resetting the counter value, confirm with the
Stop
key.
To reset the counter to zero, press the
Clr
key (F4)
The counting period is displayed in the left part of the IN window. This period is reset whenever counting is stopped.