6 Operating CCU 100
42
D52IB001EN-G
6.5.2 Controlling the Compressed-Air Cool-
ing
CCU 100 turns ON the compressed-air cooling un-
der the following conditions:
• Air cooling is turned on, if the temperature set-
point (SP) is 5 °C lower than the actual tempera-
ture of the sample holder.
• Air cooling is permanently turned on if the tem-
perature set point is below a threshold value
called Cooling Limit (CL). CL can be set
between 20 °C and 50 °C on the Parameter
Page of CCU 100.
The cooling is automatically switched OFF if
CCU 100 changes to Standby mode because of an
error and if CCU 100 is switched off completely.
TIP:
Always set the air cooling limit (CL) higher than
the temperature of the cooling water which runs
through the chamber.
TIP:
If you want to switch off the air cooling at the
end of a measurement (target temperature = 25 °C),
you can set the Cooling Limit on CCU 100 to a value
below 25 °C).
6.6 Manual Control of CCU 100
Manual operation of TTK 600 is done with the keys
on the front panel of CCU 100. Proceed as follows
to heat or cool the sample to the desired tempera-
ture:
1. Make sure that the heater is switched on.
2. Set the desired heating/cooling rate on the
parameter page of CCU 100. (See chapter
6.3.1.)
3. Push the scroll button one time (SP LED green)
and set the desired target temperature by using
the UP/DOWN buttons.
4. Wait until the displayed sample holder tempera-
ture has reached the target value.
TIP:
If you have chosen a large heating/cooling
rate, wait 1 minute before you start the X-ray scan to
make sure the sample surface has reached a con-
stant temperature.
6.7 Remote Control of CCU 100
Usually CCU 100 is integrated in the software that
controls the X-ray diffractometer. For more informa-
tion about controlling TTK 600, read the instruction
manual of the diffractometer control software.
Alternatively it is also possible to remote control the
instrument with the delivered Nambicon software
(Non-AMBIent-CONtrol) by Anton Paar GmbH. For
further information see the corresponding manual of
the Nambicon software.
The NAMBICON software is included in the delivery
of the instrument. Please follow the instructions in
the manual of the software for corresponding instal-
lation and use.