4 Installation
24
D52IB001EN-G
4.3 Mounting the Adapter on the Sam-
ple Chamber
If it is necessary to mount a motorized z-alignment
stage, see the mounting instructions delivered with
the alignment stage. The adapter to the chamber is
mounted on the z-alignment stage on delivery.
In order to mount a rigid adapter, proceed as fol-
lows:
Fig. 26: Mounting the rigid adapter stage on the
sample chamber
1. Loosen the two Allen screws (1C) and remove
the alignment equipment (1) from the base-plate
(3) of TTK 600.
2. If required, clean the O-ring on the adapter
flange and lightly coat the O-ring with vacuum
grease.
3. Use the four Allen screws (4) to mount the
adapter on the TTK 600 base-plate (3).
4. Fasten the lower part of the alignment equip-
ment (2) with the two screws (1D).
5. Fasten the two upper parts (1A) of the alignment
equipment to the TTK 600 base-plate using the
two screws (1C) in the front of the upper part.
Make sure that the alignment screw (1B) is sur-
rounded by the two parts (1A).
Adapter for Battery Sample Holders:
When changing from standard sample holders to
the battery sample holders, ensure that the guiding
pins of the TTK 600 are aligned in the correct posi-
tion on the adapter. There are different positions for
standard sample holders, the reflection battery sam-
ple holder, and the transmission battery sample
holder.
Fig. 27: Adapter for TTK 600
1 Mounting position for standard sample holders
2 Mounting position for battery sample holder reflection
3 Mounting position for battery sample holder transmis-
sion
Fig. 28: Mounting the adapter stage on the chamber for
use with different sample holders
1 TTK 600 back side
2 Guiding pins
3 Adapter for TTK 600
NOTICE
If TTK 600 has been ordered together with the
rigid adapter or motorized z-alignment stage, the
adapter/z-alignment stage is mounted on deliv-
ery.
1A
1B
1C
1D
1
2
3
4
NOTICE
The rigid height alignment tool for the rigid
adapter is optional (mat.no. 166148) and only
needed for diffractometers without a z-stage.
1
2
3
1
2
3
1
2
3
Содержание TTK 600
Страница 2: ......
Страница 74: ...Appendix B Sample Chamber Dimensions 74 D52IB001EN G Appendix B Sample Chamber Dimensions ...
Страница 75: ...Appendix C Declaration of Conformity D52IB001EN G 75 Appendix C Declaration of Conformity ...
Страница 77: ...Appendix D Spare Parts and Accessories D52IB001EN G 77 176383 COOLING MODE SWITCHING VALVE ...