C96IB01
61
EN
11.4 Exporting Adjustment Data
The currently valid adjustment data are stored on your instrument. You can transfer
these to your PC.
1.
If your PC has no IrDA interface, install the IrDA USB adapter on your PC.
2.
Press <Menu> and select
"Setup > Import / Export > Send to PC >
Adjustment"
.
3.
Hold the IrDA interface of DMA 35 to the IrDA interface of the PC to establish
communication between the two instruments.
A dialog window appears on the PC asking you to accept the file transfer.
4.
Click <Yes>.
The adjustment data are transferred and filed on the desktop as a text file.
Example:
<DMA35>
<ADJUSTMENTDATA>
<KA>1.518872E-06</KA>
<KB>1.00</KB>
<KC>1.352293E-04</KC>
<KD>-2.457641E-07</KD>
<TKA>7.068061E-04</TKA>
<TKB>2.739587E-04</TKB>
<TKC>5.822781E-08</TKC>
</ADJUSTMENTDATA>
</DMA35>
Содержание 84138
Страница 2: ......
Страница 3: ...C96IB01 3 EN DMA 35 Portable Density Meter From Firmware Version V1 79 Instruction Manual ...
Страница 72: ...72 C96IB01 EN Appendix C CE Declarations of Conformity ...
Страница 73: ...C96IB01 73 EN ...
Страница 74: ...74 C96IB01 EN Appendix D EC Type Examination Certificate ...
Страница 75: ...C96IB01 75 EN ...
Страница 76: ...76 C96IB01 EN Appendix E RFID Module Test Reports ...
Страница 77: ...C96IB01 77 EN ...
Страница 81: ...C96IB01 81 DE DMA 35 Tragbares Dichtemessgerät Ab Firmwareversion V1 79 Betriebsanleitung ...
Страница 153: ...C96IB01 153 DE Anhang C CE Konformitätserklärung ...
Страница 154: ...154 C96IB01 DE ...
Страница 155: ...C96IB01 155 DE Anhang D EG Baumusterprüfbescheinung ...
Страница 156: ...156 C96IB01 DE ...
Страница 157: ...C96IB01 157 DE Anhang E RFID Modul Testreports ...
Страница 158: ...158 C96IB01 DE ...