40
C96IB01
EN
Sample ID list
The following table explains each line of a sample ID list:
To enter a sample ID list
1.
If your PC has no IrDA interface, install the IrDA USB adapter on your PC.
2.
Hold the IrDA interface of DMA 35 to the IrDA interface of the PC to establish
communication between the two instruments.
3.
First define one or two sample IDs on the instrument, as described above. Then
press <Menu> and select
"Setup > Import / Export > Send to PC > Sample-ID
List"
.
A dialog window appears on the PC asking if you want to accept the file.
4.
Click <Yes>.
The sample ID list is transferred and filed on the desktop.
5.
Open the sample ID list on your PC in a text editor and enter further sample IDs
using the same format at will.
6.
Save your changes.
7.
Press <Menu> and select
"Setup > Import / Export > Receive from PC"
.
8.
Hold the IrDA interface of DMA 35 to the IrDA interface of the PC again to
establish communication between the two instruments.
9.
On the PC right mouse-click on the file you want to send and select
"Send to >
A nearby computer"
.
The sample ID list is sent to DMA 35 and automatically accepted.
<DMA35>
<IDLIST>
<ID>
<NAME></NAME>
Name of the first ID: <empty>
<RFID></RFID>
No RFID tag has been allocated
</ID>
<ID>
<NAME>ETHANOL</NAME>
Name of the second ID: <Ethanol>
<RFID>0100.7916.3A16</RFID>
RFID Tag 0100.7916.3A16 allocated
</ID>
<ID>
<NAME>WATER</NAME>
Name of the third ID: <Water>
<RFID></RFID>
No RFID tag has been allocated
</ID>
</IDLIST>
</DMA35>
Содержание 84138
Страница 2: ......
Страница 3: ...C96IB01 3 EN DMA 35 Portable Density Meter From Firmware Version V1 79 Instruction Manual ...
Страница 72: ...72 C96IB01 EN Appendix C CE Declarations of Conformity ...
Страница 73: ...C96IB01 73 EN ...
Страница 74: ...74 C96IB01 EN Appendix D EC Type Examination Certificate ...
Страница 75: ...C96IB01 75 EN ...
Страница 76: ...76 C96IB01 EN Appendix E RFID Module Test Reports ...
Страница 77: ...C96IB01 77 EN ...
Страница 81: ...C96IB01 81 DE DMA 35 Tragbares Dichtemessgerät Ab Firmwareversion V1 79 Betriebsanleitung ...
Страница 153: ...C96IB01 153 DE Anhang C CE Konformitätserklärung ...
Страница 154: ...154 C96IB01 DE ...
Страница 155: ...C96IB01 155 DE Anhang D EG Baumusterprüfbescheinung ...
Страница 156: ...156 C96IB01 DE ...
Страница 157: ...C96IB01 157 DE Anhang E RFID Modul Testreports ...
Страница 158: ...158 C96IB01 DE ...